KYORITSU 1062 User Manual
Page 51
109
Table 1. Input Signal for Calibration
Range
Input 1
Input 2
Unit
DC 50mV
0.000
50.000
mV
DC 500mV
500.00
-
mV
DC 2400mV
2000.0
-
mV
DC 5V
5.0000
-
V
DC 50V
50.000
-
V
DC 500V
500.00
-
V
DC 1000V
1000.0
-
V
AC 50mV*1
5.000
50.000
mV
AC 500mV
50.00
500.00
mV
AC 5V *2
0.5000
5.0000
V
AC 50V
5.000
50.000
V
AC 500V
50.00
500.00
V
AC 1000V
100.0
1000.0
V
500Ω
0.00
500.00
Ω
5kΩ
0.0000
5.0000
kΩ
50kΩ
0.000
50.000
kΩ
500kΩ
0.00
500.00
kΩ
5MΩ
0.0000
5.0000
MΩ
50MΩ
0.000
30.000
MΩ
Continuity Check ( )
0.0
500.0
Ω
DC 500 µA
0.00
500.00
µA
DC 5000 µA
0.0
5000.0
µA
DC 50mA
0.000
50.000
mA
DC 500mA
0.00
400.00
mA
DC 5A
0.0000
5.0000
A
DC 10A
0.000
10.000
A
AC 500 µA
50.00
500.00
µA
AC 5000 µA
500.0
5000.0
µA
AC 50mA
5.000
50.000
mA
AC 500mA
50.00
400.00
mA
AC 5A
0.5000
5.0000
A
AC 10A
1.000
10.000
A
*1: For model KEW1062.
*2: For model KEW1062. Calibration points added in 5 V AC range (setting, input value).
[RMS], Filter OFF, Input1 (0.50000), Input2 (5.0000)
[MEAN], Filter OFF, Input1 (0.50000), Input2 (5.0000)
[RMS], Filter ON, Input1 (0.5000) , Input2 (5.0000)
50Hz/60Hz
50Hz/60Hz