4.4 Memory Function
The instrument can save a data using with the following two types of modes.
SAVE-mode:
Saves a data for one measurement by manual operation.
LOGGING-mode: Automatically saves a data from the start of logging.
Memory capacity
SAVE-mode:
100 data
LOGGING-mode: Logging data of one time Model KEW1061 1,000 data
Model KEW1062 10,000 data
Number of saved data
Number of saved data is 4-digit numbers. When LOGGING-mode, “ L ” is attached
to the top of 4-digit numbers. The instrument allocates the smallest number, between
0000 to 9999, that has not yet been used. Use the
(RANGE) key or
(REL /%)
key switches the number of saved data.
To save a Data (SAVE-mode)
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the SAVE (HOLD) key.
(The display shows the number of saved data.)
3) Press the SAVE (HOLD) key to save the data.
Another press of the SAVE (HOLD) key saves a data for the second time measurement or
later.
4) To cancel the function, hold down the MEMORY key for one second.
(“ MEM ” symbol disappears.)
Note
HOLD data can be saved.
Hold the display and save it according to the above steps.
The number of saved data