Caution – KYORITSU 1062 User Manual
Page 33
91
4.3.12 DC, AC Current Dual Display ( •
)
1) Turn the function switch to the “µA,” “mA” or “A” position.
(If the magnitude of the current being measured is not known,
select the “A” position. Make sure the current being measured
is no more than 440 mA before the “µA” or “mA” position is
selected.)
2) Press the SELECT key three times to select DC/AC dual
display.
DC current measurement appears on the main display and AC
current on the sub-display.
3) Plug the black test lead into the “COM” input terminal and the
red test lead into the “A” input terminal.
If the current is in the order of mA or less, plug the red test lead
into the “μA/mA” input terminal.
4) Connect the test leads to the circuit under test and then read the
value when it stabilizes.
Test leads
Red
Black
Test leads
Red
Black
4.3.13 Capacitor Measurement (
)
CAUTION
To avoid damage to instrument
• Turn off the power to the circuit under test before starting measurement in order to prevent
any excessive voltage from being applied to the instrument.
• Before starting measurement, be sure to discharge the capacitor under check.
1) Turn the function switch to the “
”position.
2) Plug the test leads into the input terminals.
3) Open the test lead and press the REL key in 5nF range to adjust the
capacitance to zero. (The display shows “ 0.000 ”.)
4) Connect the test leads to the circuit under test and then read the value
when it stabilizes.
Test leads
Red
Black
Note
The value (zero adjustment) remains displayed until power-off.