KYORITSU 1062 User Manual
Page 14
72
DCV+ACV
+
KEW1061
Maximum Reading 50000, Crest factor*: <3
Range Resolution
Accuracy
Input
Impe-
dance
Maximum
Input
Voltage
DC, 10 to
20Hz
DC, 20Hz
to 1kHz
DC, 1k to
10kHz
DC, 10k
to 20kHz
5V 0.0001V 1.5+10
*1
1+10
*1
2+10
*2
11MΩ
<50pF
1000V
rms AC
1000V DC
50V
0.001V
10MΩ
<50pF
500V
0.01V
1000V*
0.1V
*2
*2
-
DCV+ACV
+
KEW1062
Maximum Reading 50000, Crest factor*: <3
Range
Reso-
lution
Accuracy
Input
Impe-
dance
Maxi-
mum
Input
Voltage
DC, 10
to 20Hz
DC,
20Hz to
1kHz
DC,
1k to
10kHz
DC,
10k to
20kHz
DC,
20k to
50kHz
DC,
50k to
100kHz
5V 0.0001V 1.5+10
*1
0.5+10
*1
1+10
*1
2+10
*2
5+20
*2
11MΩ
<50pF
1000V
rms AC
1000V
DC
50V 0.001V
10MΩ
<50pF
500V
0.01V
1000V*
0.1V
*2
*2
-
*: Crest factor <1.5 at 1000V range
Accuracy *1: At 5 to 100% of range, *2: At 10 to 100% of range
CMRR: 80dB or more DC to 60Hz (Rs=1kΩ)
Response time: Approx. 2 sec
DC Current Measurement
A
Range
Resolution
Accuracy
Voltage Drop
Maximum Input
Current
KEW1061, KEW1062
500μA
0.01μA
0.2+5
<0.11mV/μA
440mA Protected by a
440mA/1000V fuse.
5000μA
0.1μA
50mA
0.001mA
<4mV/mA
500mA*3
0.01mA
5A
0.0001A
0.6+10
<0.1V/A
10A Protected by a
10A/1000V fuse.
10A
0.001A
0.6+5
*3: Maximum measurement current : 440mA at 500mA range
Response time: 0.3 sec max.
Note : After measuring over 500mA DC (especially 10A DC), “Zero error” occurs for a
while. In that case, please wait for a while at zero input until the value stabilizes before
measuring again.