8

4.3

GFC 700 Signal converter

Versions
Integral systems (K)

GFC 700 K signal converter mounted on primary head

Remote systems (F)

GFC 700 F signal converter with wall mount (rotating design) and
additional terminal box

Option MP

signal converter equipped with magnet sensors, to set the signal
converter by means of hand held bar magnet without opening the
housing

Power supply

1. AC Version AC / DC Version

Standard

Option

1. Rated voltage

115 / 230 V

24 V AC

24 V DC

Tolerance band

+/- 13 %

20 – 27 V AC

18 – 32 V DC

Frequency

48 – 63 Hz

48 – 63 Hz

Power consumption

Max. 13 VA

approx. 10 VA

approx. 8 W

(incl. primary head)

When connected to functional extra-low voltage (24 V) safety separation (PELV) is essential

to VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or equivalent national standards.

Analog Output

• continuous flowrate measurement or measurement of sound velocity
to determine (composition of) liquid product
• all operating data configurable
• galvanically isolated
• for active and passive mode

Function

• useable as internal power supply for the binary outputs

Current

for Q = 0%: 0 – 16 mA

for Q = 100%: 4 – 20 mA

settings in 1 mA increments
(I

max

= 22 mA)

Active mode

load max. 680 Ohm

Passive mode

external voltage

≤ 18 V DC

load

≤ 680 Ohm

Low-flow cutoff

cutoff “on” value

1 – 19%

cutoff “off” value

2 – 20%

of Q

100%

,

setting in 1% increments

Time constant

0.04 – 3600 s, setting in increments of 1, 0.1 or 0.01 s

Forward / reverse measurement

direction identified via status output (or pulse output)

Internal power supply for binary outputs U

int

= 19 – 32 V DC / I

≤ 50 mA

Pulse output
Function

– continuous flow counting or measurement of sound velocity

to determine (composition of) liquid product

– all operating data configurable

– galvanically isolated

– active and passive mode

Pulse rate for Q = 100%

10 – 3600000 pulses/h

0,167 – 60000 pulses/min

0.0028 – 1000 pulses/s (= Hz)

optionally in pulse per liter, m

3

,

US gallons or user-defined unit

Pulse width

automatic: pulse duty cycle 1:1, max. 1000 Pulse/s = max. 1000 Hz

variable: 30, 50, 100, 200, 500 ms,

≤ 10 Pulse/s ≤ 10 Hz

Active mode

connection:

electronic totalizer

internal voltage:

19 – 32 V DC, from current output

load

current:

I

max

< 50 mA, operation with status

output

Passive mode

connection:

electronic or electromechanical totalizer

external

voltage:

U

ext

≤ 32 V DC / ≤ 24 V AC

load

current:

I

max

≤ 150 mA

Low-flow cutoff

cutoff “on” value: 1 – 19%

cutoff “off” value: 2 – 20%

of

Q100%

, setting in 1% increments

Time constant

0.04 – 3600 s, setting in increments of 1, 0.1 or 0.01 s

Forward / reverse measurement

direction identified via status output or current output (status output only
available in non-EEx version)