5 Installation

notes

Inlet run:

• downstream of a compressure or nozzle

≥ 40 x DN

• downstream of a fan

≥ 30 x DN

• downstream of fully open control valve

≥ 20 x DN

• downstream of a 90° bend (elbow)

≥ 20 x DN

• downstream of a reducer (

α/2 < 4°)

no additional inlet run required

Outlet run:

≥ 10 x DN (DN = meter size)

Installation
conditions

Select position such that the measuring beam is approximately horizontal.
Fit the mating flanges precisely at right angles to the pipeline.

6.1 Electrical

connection

• Power supply, power consumption and load rating
of outputs, see “Technical data”

• Current and pulse outputs (I + P) are galvanically
isolated
from all input and output circuits.

Selection of connection diagrams

Standard operation

EEx operation

Current output

2) passive/active

1) active (also standard)

Pulse output

3) passive

4) passive (also standard)

Power supply and connection terminals

5)

6)

1) Current output active

Standard and EEx



R

i

≤ 680 Ω

2) Current output passive

Standard


U

ext.

≤ 18 V DC

R

i

≤ 680 Ω

3) Pulse output active
Standard

U

Output

= 19 V

− 32 V

R1

≥ 650 Ω

I

≤ 50 mA

EC electronic totalizer

4) Pulse output passive

Standard and EEx

U

ext.

≤ 32 V DC ≤ 24 V AC

I

≤ 150 mA

EC electronic totalizer
EMC electromechanical

totalizer

5) Power supply connection and all outputs
standard


L N

L~ L~

only internal protective conductor

connection terminal

6) Power supply connection and all outputs

EEx version



L N

L~

L~


power supply

protective conductor

terminal



6.2 Electrical installation to the mains supply voltage

The GFM 700 is intended for permanent connection to the mains. It is required (for example for service) to mount an
external switch or circuitbreaker near the product for disconnection from the mains. It must be reached easily by the
operator and marked as the disconnecting device for this product. The switch or circuitbreaker has to be suitable for the
application and shall also be in accordance with local (safety) requirements and requirements of the building installation.

The protective conductor clamp terminal size M5, press-fitted, in the terminal compartment (Inear the power connection
terminals) shall always be connected to the protective earth conductor of the main supply. Conductors up to 4mm

2

can be

connected.

The diameter of the conductors of the main supply, including the protective earth conductor shall to be in accordance with
the general requirements.

binary

outputs

current

output

binary
oututs


current
output