KROHNE GFM 700 EN User Manual
Page 13
13
Fct. Text
Description and settings
3.0
INSTALL
Main menu 3.0 Installation
3.1.0
BASIS.PARAM.
Submenu 3.1.0 Basisparameters
3.1.1
FULL SCALE
Full-scale range for flowrate Q
100%
Unit:
select from list
under Fct. 3.2.1
Value: 9.5*10
-7
- 150.8 m
3
/Sec or
3.9 - 1,987,200 US Gal/min
After selecting unit, call numerical
value with
↵ key, 1
st
digit flashes.
3.1.2
REV. SCALE
Different range for reverse flow
required? Setting NO or YES
3.1.3 REV.
VALUE Full-scale range for reverse flow
(appears only if YES set
under Fct. 3.2.1)
Unit:
select from list
under Fct. 3.2.1
Value: 9.5*10
-7
- 150.8 m
3
/Sec or
3.9 - 1,987,200 US Gal/min
Value must not be larger
than that of Fct. 3.1.1!
After selecting unit, call numerical
value with
↵ key, 1
st
digit flashes.
3.1.4
ZERO SET
Zero setting
• FIXED.VALUE
• VALUE.MEASU.
(Carry out only at “zero” flow and
with completely filled measuring
tube).
1) Inquiry: CALIB. NO or YES
2) if YES: calibration (duration
approx. 20 Sec) with zero display
in PERCENT of Q
100%
3) Inquiry: STORE NO or YES.
3.1.5
METER SIZE
Meter size
Unit:
mm or inches
Value: 25 - 1200 mm or
0.98 - 48 inches
After selecting unit, call numerical
value with
↵ key, 1
st
digit flashes.
3.1.6
GK VALUE
Primary head constant GK
(see also primary head nameplate).
Range: 0.5 - 14
3.1.7
FLOW DIR
Define direction of forward flow,
see Sect. 5.4.
Setting: + or
−
, acc. to direction of
arrow
on primary head.
3.1.8
MIN SP. S.
Minimum sound velocity
Minimum value used for I
0%
or F
0%
(when function SOUND.VELO.
selected in 3.3.1 or 3.4.1)
Value: 0 to 5000 m/s
3.1.9
MAX SP. S
Maximum sound velocity
Maximum value used for I
100%
or F
100%
(when function SOUND.VELO.
selected in 3.3.1 or 3.4.1)
Value: 1 - 5000 m/s