Fct. Text

Description

and settings

3.3.0

CUR.OUTP. I Submenu 3.3.0 Current output I

3.3.1 FUNCTION

I

Function, current output I

OFF (switched off)

F/R IND.

(F/R indication, e.g. for F)

1 DIR. (1 flow direction)

I

(Forward / Reverse flow,

e.g. in 0 - 20 mA range:

F=10 - 20 mA and R=10 - 0 mA)

2 DIR. (Forward/Reverse flow,

F/R-measurement)

SP. SOUND (sound velocity)

3.3.2

RANGE I

Range for current output I, see

Sect. 5.7.2

0 - 20 mA

4 - 20 mA

OTHER RANGE

(see Fct. 3.3.3, 3.3.4 + 3.3.5)

3.3.3

I 0 PCT.

Current for 0% flow (I

0%

)

(appears only if OTHER RANGE

set under Fct. 3.3.2).

Value: 00 to 16 mA

3.3.4

I 100 PCT.

Current for 100% flow (I

100%

)

of full-scale range (Fct. 3.1.1)

(appears only if OTHER RANGE

set under Fct. 3.3.2).

Value: 04 - 20 mA

(value must be at least 4 mA

greater than that of Fct. 3.3.4).

3.3.5

I MAX mA

Current limitation (I

max

)

see Fct. 5.7.2

(appears only if OTHER RANGE is

set under Fct. 3.3.2)

Value: 04 20 mA

(value must be at least 4 mA

greater than that of Fct. 3.3.4).

3.3.6 TIMECONST.

I

Time constant of current output I

Value: 0.04 - 3600 Sec

3.3.7 L.F.CUTOFF

I

Low flow cutoff (SMU) for current

output required?

Setting: NO or YES

3.3.8 L.F.CUTOFF

ON

Cutoff “on” value for SMU-I

(appears only if YES set

under Fct. 3.3.7)

Value: 01 - 19 PERCENT of Q

100%

(Fct. 3.1.1)

3.3.9 L.F.CUTOFF

OFF

Cutoff “off” value for SMU-I

(appears only if YES set

under Fct. 3.3.7)

Value: 02 - 20 PERCENT of Q

100%

(Fct. 3.1.1), value must be greater

than that of Fct. 3.3.8.