Operation of the signal converter

7.1

Table of settable functions

Fct. Text

Description and settings

1.0

OPERATION

Main menu 1.0 Operation

1.1.0 BASIC.

PARAM

Submenu 1.1.0 Basis parameters

1.1.1 FULL SCALE

Full-scale range for flowrate Q

100%

,

see Fct. 3.1.1

1.1.2 REV. SCALE

Different range for reverse flow

required? See Fct. 3.1.2

1.1.3 REV.

VALUE Full-scale range for reverse flow

Q

R100%

, see Fct. 3.1.3

1.1.4 ZERO SET.

Zero setting, see Fct. 3.1.4

1.2.0 DISPLAY

Submenu 1.2.0 Display

1.2.1 DISP.

FLOW Unit

for flowrate display,

see Fct. 3.2.1

1.2.2 DISP.

TOTAL. Function of totalizer display,

see Fct. 3.2.2

1.2.3 UNIT TOTAL.

Unit for totalizer display required?

See Fct. 3.2.3

1.2.4 DISP. SP. S.

Display of sound velocity required?

See Fct. 3.2.4

1.2.5 CYCL. DISP.

Cyclic display required?

See Fct. 3.2.5

1.3.0 CUR.

OUTP. I

Submenu 1.3.0 Current output I

1.3.1 TIMECONST.I Time

constant of current output I,

see Fct. 3.3.6

1.3.2 L.F.CUTOFF I

Low-flow cutoff (SMU) for current

output required? See Fct. 3.3.7

1.3.3 CUTOFF

ON

Cutoff “on” value SMU-I,

see Fct. 3.3.8

1.3.4 CUTOFF OFF

Cutoff “off” value SMU-I,

see Fct. 3.3.9

1.4.0 FREQ. OUTP.F Submenu 1.4.0

Frequency output F

1.4.1 PULSRATE

or

PULSE/UNIT

Pulse rate for 100% flowrate or for

sound velocity, see Fct. 3.4.3 or

Pulse value for flowrate unit,

see Fct. 3.4.3

1.4.2 L.F.CUTOFF F Low-flow cutoff (SMU) for frequency

output required? See Fct. 3.4.3

1.4.3 CUTOFF ON

Cutoff “on” value SMU-F,

see Fct. 3.4.7

1.4.4 CUTOFF OFF

Cutoff “off” value SMU-F,

see Fct. 3.4.8

2.0

TEST

Main menu 2.0 Test functions

2.1

TEST DISP.

Carry out display test (Sect. 7.1.1)

Start with

→ key, duration approx. 30

Sec. Test interruption with

↵ key.

2.2

TEST I

Test current output I (Sect. 7.1.2)

0 mA

4 mA

20 mA

2 mA

10 mA

22 mA

Displayed value present directly at

current output. Actual value again

present at output after pressing the

key.

2.3

TEST F

Test frequency output F (Sect. 7.1.3)

1 Hz

100 Hz

10 Hz

• 1000 Hz

Displayed value present directly at

current output. Actual value again

present at output after pressing the

key.

2.4

PROCESSOR

Test microprocessor (Sect. 7.1.4)

Start with

↵ key, duration approx. 2

Sec. End of test: NO ERROR or

ERROR displayed.