KROHNE GFM 700 EN User Manual
Page 18
18
Fct. Text
Description
and settings
3.5.9 TOTAL.
RESET
Totalizer reset (+ and −- totalizer
together)
Inquiry: NO or YES
3.5.10 ENABL.
RESET
Enable totalizer reset
for RESET/QUIT menu.
Inquiry: NO or YES
3.5.11 PLAUSIB
ERR.
Error limit in % of measured value for
plausibility statement. Measured
values that are outside the specified
band are not processed. Every
measured value outside the specified
band will increase an internal counter
by “1”, until a maximum counter
value (see Fct. 3.6.13) has been
reached. The corresponding
measurement channel will then be
made inactive and an indication will
be visible on the display.
Value setting: 1 - 99 PERCENT
Factory setting: 20 PERCENT
3.5.12 WEIGHT
P.OK
Weight factor for correct
measurements. The internal
plausibility counter is increased by
the number programmed, when the
measured value is correct. The
higher the number the faster an
inactive channel will become active
again.
Value setting: 1 - 50
Factory setting: 4
3.5.13 N.ER.PLAU
SIB.
Limit value for the counter of
incorrectly measurements (see Fct.
3.6.11). When ‘0’ is set, the
plausibility function will become
inactive.
Value setting: 0 - 10,000
Factory setting: 0