beautypg.com

2 error messages – KROHNE GFM 700 EN User Manual

Page 19

background image

19

7.2 Error

messages

Fct. Text

Description

and settings

4.0

PARAM.

ERROR

Main menu 4.0 Parameter error

4.1.0

FLOW

VELOC.

FLOW VELOCITY “v” incorrect:

Ensure condition

0.5 m/s

v 60 m/s

or 1.5 ft/s

v 180 ft/s is met!

4.1.1

FULL SCALE Full-scale range for flowrate Q

100%

,

see Fct. 3.1.1

4.1.2

METER SIZE Meter size, see Fct. 3.1.5

4.2.0

F/R FLOW

FULL-SCALE RANGE(S) for

forward/reverse flow incorrect:

Ensure condition F

R is met!

4.2.1

FULL SCALE Full-scale range for forward

flowrate Q

F100%

, see Fct. 3.1.1

4.2.2

REV. SCALE Different range for reverse flow

required? See Fct. 3.1.2

4.2.3 REV.

VALUE

Full-scale range for reverse flow

Q

R100%

, see Fct. 3.1.3

4.3.0

I RANGE

CURRENT OUTPUT I RANGE

incorrect:

Ensure condition

I

100%

I

0%

4 mA is met!

4.3.1

I 0 PCT

Current for 0% flow (I

0%

), see Fct.

3.3.3

4.3.2

I 100 PCT

Current for 100% flow (I

100%

), see

Fct. 3.3.4

4.4.0

I MAXIMUM

CURRENT LIMITATION

incorrect:

Ensure condition

I

max

I

100%

is met!

4.4.1

I 100 PCT

Current for 100% flow (I

100%

), see

Fct. 3.3.4

4.4.2

I MAX mA

Setting of max. output current

(I

max

),

see Fct. 3.3.5

4.5.0

LFC. I RANG. LOW-FLOW CUTOFF RANGE I

incorrect:

Ensure condition

cutoff “off”.

Cutoff “on”

1% is met!

4.5.1 L.F.

CUTOFF

I

Low-flow cutoff (SMU) for current

output required? See Fct. 3.3.7

4.5.2

CUTOFF ON Cutoff “on” value SMU-I, see Fct.

3.3.8

4.5.3 CUTOFF

OFF

Cutoff “off” value SMU-I, see Fct.

3.3.9

4.6.0

LFC. F

RANG.

LOW-FLOW CUTOFF RANGE F

incorrect:

Ensure condition cutoff “off”.

Cutoff “on”

1% is met!

4.6.1 L.F.

CUTOFF

F

Low-flow cutoff (SMU) for

frequency output required?

See Fct. 3.4.6

4.6.2

CUTOFF ON Cutoff “on” value SMU-F,

see Fct. 3.4.7

4.6.3 CUTOFF

OFF

Cutoff “off” value SMU-F,

see Fct. 3.4.8