2 error messages – KROHNE GFM 700 EN User Manual
Page 19
19
7.2 Error
messages
Fct. Text
Description
and settings
4.0
PARAM.
ERROR
Main menu 4.0 Parameter error
4.1.0
FLOW
VELOC.
FLOW VELOCITY “v” incorrect:
Ensure condition
0.5 m/s
≤ v ≤ 60 m/s
or 1.5 ft/s
≤ v ≤ 180 ft/s is met!
4.1.1
FULL SCALE Full-scale range for flowrate Q
100%
,
see Fct. 3.1.1
4.1.2
METER SIZE Meter size, see Fct. 3.1.5
4.2.0
F/R FLOW
FULL-SCALE RANGE(S) for
forward/reverse flow incorrect:
Ensure condition F
≥ R is met!
4.2.1
FULL SCALE Full-scale range for forward
flowrate Q
F100%
, see Fct. 3.1.1
4.2.2
REV. SCALE Different range for reverse flow
required? See Fct. 3.1.2
4.2.3 REV.
VALUE
Full-scale range for reverse flow
Q
R100%
, see Fct. 3.1.3
4.3.0
I RANGE
CURRENT OUTPUT I RANGE
incorrect:
Ensure condition
I
100%
−
I
0%
≥ 4 mA is met!
4.3.1
I 0 PCT
Current for 0% flow (I
0%
), see Fct.
3.3.3
4.3.2
I 100 PCT
Current for 100% flow (I
100%
), see
Fct. 3.3.4
4.4.0
I MAXIMUM
CURRENT LIMITATION
incorrect:
Ensure condition
I
max
≥ I
100%
is met!
4.4.1
I 100 PCT
Current for 100% flow (I
100%
), see
Fct. 3.3.4
4.4.2
I MAX mA
Setting of max. output current
(I
max
),
see Fct. 3.3.5
4.5.0
LFC. I RANG. LOW-FLOW CUTOFF RANGE I
incorrect:
Ensure condition
cutoff “off”.
Cutoff “on”
≥ 1% is met!
4.5.1 L.F.
CUTOFF
I
Low-flow cutoff (SMU) for current
output required? See Fct. 3.3.7
4.5.2
CUTOFF ON Cutoff “on” value SMU-I, see Fct.
3.3.8
4.5.3 CUTOFF
OFF
Cutoff “off” value SMU-I, see Fct.
3.3.9
4.6.0
LFC. F
RANG.
LOW-FLOW CUTOFF RANGE F
incorrect:
Ensure condition cutoff “off”.
Cutoff “on”
≥ 1% is met!
4.6.1 L.F.
CUTOFF
F
Low-flow cutoff (SMU) for
frequency output required?
See Fct. 3.4.6
4.6.2
CUTOFF ON Cutoff “on” value SMU-F,
see Fct. 3.4.7
4.6.3 CUTOFF
OFF
Cutoff “off” value SMU-F,
see Fct. 3.4.8