2 error limits – KROHNE IFC 110 Converter EN User Manual
Page 40
40
IFC 110 F
05/2003
10.2 Error limits
Display, digital values, pulse output
F maximum error in % of measured value (MV), not typical values
v Flow velocity in m/s and ft/s
Reference conditions similar to EN 29 104
Product
water at 10 – 30°C/ 50 – 86°F
Electrical conductivity
> 300 µS/cm
Power supply (rated voltage)
U
N
(± 2%)
Ambient temperature
20 – 22°C / 68-71.6 °F
Warm-up time
60 min
Max. calibration equipment error
10
× smaller than F
Inlet / outlet runs
10
× DN / 2 × DN (DN = meter size)
Flow sensor
properly grounded and centered
Calibrated on EN 17025 accredit to calibration equipment in direct volumetric comparsion.
*
VARIFLUX 6000 F (DN 2.5 – 4
and 1/10’’ – 1/6’’)
additional error ± 0,3% of MV
MV Measured
Value
z
= 1 mm/s = 0.04 inch/s
Flow Meter
size
Standard details
Option (extra charge)
sensor
DN mm
inch
v
≥ 1.0 m/s
v < 1.0 m/s
Curve
v
≥ 1.0 m/s
v < 1.0 m/s
Curve
v
≥ 3.3 ft/s
v < 3.3 ft/s
v
≥ 3.3 ft/s
v < 3.3 ft/s
VARIFLUX
2.5- 6
1
/
10
-
1
/
4
± 0.5% of MV
± 0.4% of MV + z
C
– –
–
6000 F
10 - 80
3
/
8
- 3
± 0.3% of MV
± 0.2% of MV + z
B
– –
–
PROFIFLUX
2.5- 6
1
/
10
-
1
/
4
± 0.5% of MV
± 0.4% of MV + z
C
– –
–
5000 F
10 - 100
3
/
8
- 4
± 0.3% of MV
± 0.2% of MV + z
B
± 0.2% of MV
± 0.1% of MV + z
A
ALTOFLUX
10. - 25
3
/
8
- 1
–
–
4000 F
32 -1600
1
1
/
4
-64
± 0.3% of MV
± 0.2% of MV + z
B
± 0.2% of MV
± 0.1% of MV + z
A
ALTOFLUX
2000 F
150 - 250
6 -10
± 0.3% of MV
± 0.2% of MV + z
B
± 0.2% of MV
± 0.1% of MV + z
A
ECOFLUX
1000 F
10 - 150
3
/
8
- 6
± 0.5% of MV
± 0.4% of MV + z
B
– –
–
M 900
10. - 25
3
/
8
- 1
–
–
–
32 - 300
1
1
/
4
-12
± 0.3% of MV
± 0.2% of MV + z
B
± 0.2% of MV
± 0.1% of MV + z
A
Current output
same error limits as above, additionally ± 10
µA
Reproducibility and
repeatability
0,1% of MV, minimum 1 mm/s / 0.04 inch/s at constant flow
External influences
typical values maximum
values
Ambient temperature
Pulse output
0,003% of MV (1) 0,01 % of MV (1)
per 1 K / 1.8° F
Current output
0,01 % of MV (1) 0,025% of MV (1)
}
temperature variation
Power supply
< 0,02 % of MV
0,05 % of MV
at 10% variation
Load
< 0,01 % of MV
0,02 % of MV, at max. load, see Sect. 10.1
(1)
All KROHNE signal converters undergo burn-in tests, duration minimum 20 hours at varying ambient temperatures
– 20 to + 60°C/– 4 to + 140°F. The tests are controlled by computers.