2 error limits – KROHNE IFC 110 Converter EN User Manual

Page 40

background image

40

IFC 110 F

05/2003

10.2 Error limits

Display, digital values, pulse output
F
maximum error in % of measured value (MV), not typical values
v Flow velocity in m/s and ft/s

Reference conditions similar to EN 29 104

Product

water at 10 – 30°C/ 50 – 86°F

Electrical conductivity

> 300 µS/cm

Power supply (rated voltage)

U

N

(± 2%)

Ambient temperature

20 – 22°C / 68-71.6 °F

Warm-up time

60 min

Max. calibration equipment error

10

× smaller than F

Inlet / outlet runs

10

× DN / 2 × DN (DN = meter size)

Flow sensor

properly grounded and centered

Calibrated on EN 17025 accredit to calibration equipment in direct volumetric comparsion.








*

VARIFLUX 6000 F (DN 2.5 – 4

and 1/10’’ – 1/6’’)

additional error ± 0,3% of MV


MV Measured

Value


z

= 1 mm/s = 0.04 inch/s

Flow Meter

size

Standard details

Option (extra charge)

sensor

DN mm

inch

v

≥ 1.0 m/s

v < 1.0 m/s

Curve

v

≥ 1.0 m/s

v < 1.0 m/s

Curve

v

≥ 3.3 ft/s

v < 3.3 ft/s

v

≥ 3.3 ft/s

v < 3.3 ft/s

VARIFLUX

2.5- 6

1

/

10

-

1

/

4

± 0.5% of MV

± 0.4% of MV + z

C

– –

6000 F

10 - 80

3

/

8

- 3

± 0.3% of MV

± 0.2% of MV + z

B

– –

PROFIFLUX

2.5- 6

1

/

10

-

1

/

4

± 0.5% of MV

± 0.4% of MV + z

C

– –

5000 F

10 - 100

3

/

8

- 4

± 0.3% of MV

± 0.2% of MV + z

B

± 0.2% of MV

± 0.1% of MV + z

A

ALTOFLUX

10. - 25

3

/

8

- 1

4000 F

32 -1600

1

1

/

4

-64

± 0.3% of MV

± 0.2% of MV + z

B

± 0.2% of MV

± 0.1% of MV + z

A

ALTOFLUX
2000 F

150 - 250

6 -10

± 0.3% of MV

± 0.2% of MV + z

B

± 0.2% of MV

± 0.1% of MV + z

A

ECOFLUX
1000 F

10 - 150

3

/

8

- 6

± 0.5% of MV

± 0.4% of MV + z

B

– –

M 900

10. - 25

3

/

8

- 1

32 - 300

1

1

/

4

-12

± 0.3% of MV

± 0.2% of MV + z

B

± 0.2% of MV

± 0.1% of MV + z

A

Current output

same error limits as above, additionally ± 10

µA

Reproducibility and

repeatability

0,1% of MV, minimum 1 mm/s / 0.04 inch/s at constant flow

External influences

typical values maximum

values

Ambient temperature

Pulse output

0,003% of MV (1) 0,01 % of MV (1)

per 1 K / 1.8° F

Current output

0,01 % of MV (1) 0,025% of MV (1)

}

temperature variation

Power supply

< 0,02 % of MV

0,05 % of MV

at 10% variation

Load

< 0,01 % of MV

0,02 % of MV, at max. load, see Sect. 10.1

(1)

All KROHNE signal converters undergo burn-in tests, duration minimum 20 hours at varying ambient temperatures
– 20 to + 60°C/– 4 to + 140°F. The tests are controlled by computers.