3 Start-up


• Before connecting to power, check that the instrument is correctly installed as described in

Sections 1 and 2.


• The flowmeter, flow sensor and signal converter are delivered ready for operation. All

operating data are set at the factory in accordance with your specifications.
Please also refer to Section 2.7 ”Standard factory settings”.


• Switch on the power supply. The flowmeter immediately begins to measure the flow.


• When the power supply is switched on, the display successively shows START UP and

READY. Then the current flow rate and/or the current totalizer count are displayed. Displays
are either steady or cyclic depending on the setting described for Fct. 1.04.


2 light-emitting diodes (LED) in the ”diagnostics” field on the front panel of the signal

converter indicate the status of measurement.



LED displays

Status of measurement

Green ”normal” LED

Everything O.K.

is flashing

Green ”normal” LED

Momentary overload of outputs and/or A/D converter.

and red ”error” LED

Detailed error messages by setting Fct. 1.04 DISPLAY,

are flashing alternately

subfunction ”MESSAGES” to ”YES”, see Sections 4.4 and 5.5.

Red ”error” LED is

Fatal Error, see Sections 7.3 and 7.4

flashing


IMPORTANT!
For EEx versions, also pay regard to all directions included in Sect. 6.1 and 13.
Only the EEx flow sensor may be installed in the hazardous area. The EEx
certified signal converter must be installed outside the hazardous area!