beautypg.com

KROHNE IFC 110 Converter EN User Manual

Page 39

background image

05/2003

IFC 110 F

39

Status outputs (passive)

D1 / D2 / A2

A1 (can also be operated as pulse output)

Function, set for

trip point

trip point

flow direction

flow direction

automatic

range

change

automatic range change

error identification

error identification

overdriving

overdriving

empty pipeline

empty pipeline

Terminals

D1 / D2 / D

⊥ / A2 / A ⊥

A1 / A

Please note:

D

⊥ common reference potential for D1 and D2

A

⊥ common reference potential for A1 and A2

Electrical data

galvanically isolated

galvanically isolated, but not from A2

U

≤ 32 V DC / ≤ 24 V AC

U

≤ 32 V DC / ≤ 24 V AC

I

≤ 100 mA, any polarity

I

≤ 100 mA, any polarity

or U

≤ 32 V DC, I ≤ 200 mA, note polarity

Control inputs C1 and C2 (passive)

Function, set for

automatic range change, totalizer reset, error reset, start self-test,

set outputs to min. values or hold last measured values of outputs

Terminals

C1 / C

⊥ and C2 / C ⊥, galvanically isolated

Please note: C

⊥ common reference potential for C1 and C2

U = 8 – 32 V DC, I

≤ 10 mA, any polarity

Internal power supply

for passive outputs and inputs and external receiver instruments

Terminals

E + and E –, please note polarity, galvanically isolated

Electrical data

U = 24 V DC / R

i

= approx. 15 Ohm / I

≤ 100 mA

Time constant

0.2 – 99.9 s, adjustable in increments of 0.1 second

Low-flow cutoff

cutoff ”on“ value: 1 – 19 %

cutoff ”off“ value: 2 – 20 %

}

of Q

100%

, adjustable in 1% increments

Local display and operation

3-line back-lit LCD

Display function

actual flowrate, forward, reverse, sum totalizers (7 digits)

or 25-character bar graph with percent display and status messages

Units:

actual flowrate

m³/h, liter/s., US gallons/min or user-defined unit, e. g. hecto liter/h

totalizer

m³, liter, or US gallons or user-defined unit (adjustable counting time till overflow)

Language of plain texts

English, German, French, Swedish, others on request

Display:

1st line

8-character, 7-segment, numerical and sign display,

and symbols for key acknowledgement

2nd line

10-character, 14-segment, text display

3rd line

6 markers to identify display in measuring mode

Operation elements

15 keys or as option with 3 additionally magnetic sensors for operation without
opening the housing

Electrode circuit

Type of protection

intrinsic safety [EEx ib IIC]

Max. values (cumulative)

U

0

= 18 V / I

0

= 40 mA / P

0

= 80 mW

Kinked characteristic

capacitance C

0

≤ 225 nF / inductance L

0

≤ 5 mH

Field power supply

Type

pulsed bipolar DC field for all KROHNE primary heads,

galvanically isolated from all input and output circuits

Terminals

2 x 7 and 8

Current / voltage

± 0.125 A (± 5%) / U

N

≤ 40 V DC (frequency controlled)

Clock frequency

1

/

36

to ½ of power frequency,

configurable to the calibration data of the primary head

Internal fuse protection

I

N

≤ 160 mA

Power supply

AC version

AC / DC version (switch-selectable)

standard

option, in preparation

Voltage range (without change over)

100 – 230 V AC

24 V AC

24 V DC

Tolerance band

85 – 255 VAC

20.4 – 26.4 V AC

18 – 31.2 V DC

Safety value

U

m

= 253 V

U

m

= 253 V

U

m

= 253 V

Frequency

48 – 63 Hz

48 – 63 Hz

Power consumption (incl. primary head)

18 VA, typical (max. 25 VA) 18 VA, typical (max. 25 VA) 18 W, typical (max. 18 W)

When connected to a functional extra-low voltage, 24 V AC / DC, protective separation (PELV) must be
ensured (VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or equivalent national standards).

Approvals and housing

Material of field housing

die-cast aluminium with polyurethane finish

Ambient temperature

• operation

standard

-25 to +60 °C / -13 to +140 °F

EEx

-20 to +55 °C / - 4 to +131 °F

EEx, special “S”

-40 to +55 °C / -40 to +131 °F

• storage

all versions

-40 to +60 °C / -40 to +140 °F

Protection category (IEC 529 / EN 60 529) IP 65, equivalent to NEMA 4 / 4X
EU_EMC Directives

to EN 61326-1 (1977) and A1 (1998) directives and NAMUR Standard NE 21

Certificates and approvals

II (2) G [EEx ib] IIC
PTB 02 ATEX 2136 X