1.3.5 Connection diagrams for power supply and flow sensors 
 
Important remarks for circuit diagrams PLEASE NOTE ! 
 
• The figures in brackets indicate the stranded drain wires of the shields (see cross-sectional 
drawings of signal lines in Section 1.3.1).
 
• Electrical connection to VDE 0100 ”Regulations for the erection of power installations with 
nominal voltages up to 1000 V”
 
• Power supply 24 V AC / DC: 
protective extra-low voltages (PELV) acc. to VDE 0100/ 
VDE 0106 and/or IEC 364/IEC 365, or corresponding 
national regulations. 
 
• Systems to be used in hazardous areas are subject to special regulations applying to 
electrical connections (see Section 1.3.6) for hazardous-duty instruments.
 
• PE = protective conductor 
FE = functional ground conductor
 
 
IMPORTANT! 
 
Electrical connection of EEx flow sensors and EEx signal converters to be carried 
out as described in Sect. 1.3.6. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
*
Do not remove the internal connection (cable) inside the terminal compartment of the signal 
converter (yellow/green wire) between the U-clamp terminal and terminal 10 (protective 
conductor for protection class I instruments). 