beautypg.com

KROHNE IFC 110 Converter EN User Manual

Page 31

background image

05/2003

IFC 110 F

31

Fct. Text

Description

and

setting

1.07 STATUS

A1

Status

output

A1

(terminal A1 connected as status output A1 or

as a 2nd pulse output A1, see Fct. 3.07 HARDWARE, ”terminal A1”)

1.08 STATUS

A2

Status

output

A2

1.09

STATUS D1

Status output D1

1.10

STATUS D2

Status output D2

● OFF

● ON

● ALL ERROR

● FATAL.ERROR

● INVERS D1 (inverse mode of D1 and D2)

● INVERS A1 (inverse mode of A1 and A2 possible only if A1

is operated as status output, see Fct. 3.07 HARDWARE, ”terminal A1”)

● SIGN I, P or P2

(F/R

mode)

● OVERFL. I, P or P2

(overloading

the

outputs)

}

dynamic behaviour
of outputs see Fct. 1.02
TIMECONST.:
I = ONLY I
P or P2 = ALL

● EMPTY PIPE (”tube empty” signal only with built-in option)

● TRIP. POINT

Presskey to change to character.

Selection:

● + DIR.

● - DIR.

● 2 DIR.

Presskey to change to numerical setting.

Setting

range: 000 - 150 PERCENT

● AUTO. RNG. Setting range: 05-80 PERCENT (= lower to upper range ratio

1:20 to 1:1.25, value must be higher than that of Fct. 1.03 L.F. CUTOFF)

Fct. 1.07 to 1.1

0

are configured in

accordance

w

ith the same setting

mode. Functions

set for one of th

e

status outputs ar

e no longer available

for the oth

er stat

us outputs.

Presskey to change to numerical setting.

Presskey to return to Fct. 1.06, 1.07, 1.08 or 1.09.

Fct.

Text

Description and setting

1.11

CONTROL C1

Control input C1 and C2

1.12 CONTROL

C2 ● OFF

● EXT. RNG. (external range change)

Setting

range: 05-80 PERCENT (= lower to upper range ratio 1:20 to 1:1.25,

value must be higher than that of Fct. 1.03 L.F. CUTOFF)

Presskey to change to numerical setting.

● OUTP. HOLD (hold output values)

● OUTP. ZERO (set outputs to ”min. values”)

● TOTAL.RESET (reset the totalizer)

● ERROR.RESET (delete error messages)

Press key to return to Fct. 1.11 or 1.12 CONTROL C1 or C2

Fct.

Text

Description and setting

2.0

TEST

Test menu

TEST Q

Test measuring range Q

Precautionary

query

● SURE NO

Press

key to return to Fct. 2.01 “TEST Q”.

● SURE YES

Press

key, then usekey to

select value: -110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.

of set full-scale range Q

100%

.

Displayed value is available at outputs I and P.

Press

key to return to Fct. 2.01 “TEST Q”.

2.02

HARDW. INFO

Hardware information and error status

Before consulting factory, please note down all 6 codes.

MODUL ADC

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

key to transfer to “MODUL IO“.

MODUL IO

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

key to transfer to “MODUL DISP.“.

MODUL DISP.

X . X X X X X . X X

Press

key to return to

Y Y Y Y Y Y Y Y Y Y

Fct. 2.02 “HARDW. INFO“.

2.03

HARDW. TEST

Hardware test (Precautionary query)

- SURE NO

Press

key to return to Fct. 2.03 ”HARDW. TEST”.

-

SURE

YES Press key to start test, duration approx. 60 s

If errors are found, the first one is displayed. Press

↓ key to

display next error. List of errors see Section 4.5.

Press

key to return to Fct. 2.03 ”HARDW. TEST”.