7 examples of signal converter settings – KROHNE IFC 110 Converter EN User Manual
Page 35
05/2003
IFC 110 F
35
4.6
Resetting the totalizer and deleting error messages, RESET/QUIT menu
Delete error messages in RESET/QUIT menu
Key
Display Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
CodE 2
- -
Key-in entry code 2 for RESET/QUIT menu:
↑ →
↑ →
ERROR QUIT.
Menu for error acknowledgement
→
QUIT. NO
Do not delete error messages,
press
↵ twice to return to measuring mode.
↑
QUIT. YES
Delete error messages
↵
ERROR QUIT.
Error messages deleted.
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
Reset totalizer in RESET/QUIT menu
Key
Display Description
- - - - - - -
- - - - - / - - -
Measuring mode
CodE 2
- -
Key-in entry code 2 for RESET/QUIT menu:
↑ →
↑ →
ERROR QUIT.
Menu for error acknowledgement
↑
TOTAL.RESET
Menu for resetting totalizer
→
RESET NO
Do not reset totalizer,
press
↵ twice to return to measuring mode
↑
RESET YES
Reset totalizer
↵
TOTAL.RESET
Totalizer is reset
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
4.7
Examples of signal converter settings
In the following example the cursor or flashing part of the display is shown in bold characters.
• Change measuring range of current output and value for error messages (Fct. 1.05):
• Change measuring range from 04-20 mA to 00-20 mA
• Change value for error messages from 0 mA to 22 mA
Key Display
Description
→
If ”YES” is selected in Fct. 3.04 ENTRY
CODE, enter the 9-digit entry CODE 1:
→ → → ↑ ↑ ↑ ↵ ↵ ↵
Fct.
1.00
OPERATION
→
Fct. 1.01
FULL SCALE
4x
↑
Fct. 1.05
CURRENT I
→
FUNCT.
I
→ ↵
RANGE
I
→
04-20
mA
old current range
2x
↑
00-20
mA
new current range
↵
I
ERROR
→
0
mA
old value for error messages
↑
22
mA
new value for error messages
↵
Fct. 1.05
CURRENT I
↵
Fct. 1.00
OPERATION
↵
STORE YES
↵
- - - - - - -
- - - - - / - - -
Measuring mode with new current output data