Installation and Operating Instructions BM 26 A

1.2

BM 26 A construction details


The diagram below shows a standard BM 26 A/C/RR without level transducer and switch options

1 Vent

plug

2 Measuring

tube

3

Red triangle reference mark giving
factory-calculated zero position of
measuring scale

4

Glass tube for indicator

5 Scale

6

Float (magnets mounted in lower half)

7 Liquid

level

8

Mid-point of indicator

9

Indicator (standard follower bar)**

10

Zero point of measurement
(centreline of lower connecting flange)

11 Scale

zero

12

Stop (red KROHNE plate)

13 Instrument

nameplate

* optional flap indicator available (see section 8)


Details:
Item 12 – The red KROHNE plate at the bottom of the scale (item 5) has a label listing the product
and the operating conditions. For example :

Working pressure :
Working temperature :
Density :
Fluid :

6 bar g
25 C
0.778 s g.
cyclohexane