Location

Do not expose the compact flowmeter to direct sunlight Install a

sunshade if necessary.

Do not expose to intense vibration. If necessary support the pipeline

to the left right of the flowmeter.

The rotating design of the housing makes it easier to connect the two

cables for power and outputs to the terminals in the rear terminal
box.

Cable diameter

To conform to protection category requirements, observe

the following recommendations :

Cable diameter: 8 to 13 mm (0.31" to 0.51")

Enlarge the inside diameter of the Screwed conduit entry by

removing the appropriate onion ring(s) from the seal, only if cables
have extremely tight fit.

Fit blanking plug PG16 and apply sealant to unused cable entries.

Do not kink cables at conduit entries

Provide water drip point (U bend in cable).

Conduit Installation, general wiring considerations

When electrical codes require conduit, it must be installed in such a
manner that the meter connection compartment remains at all dry
times.

In case of functional extra low voltages (24VDC) protective

separation in conformity with VDE 400 part 410, or equivalent
national standard, must be ensured.

Hazardous area application will require the connection of

protective Earth.

Connection to power, VFM 1091 K

Current and frequency outputs are galvanically isolated when the

optional ground above is not used and can therefore be
simultaneously connected to a receiver instrument which is
grounded or separately connected to two receiver instruments.

Ensure that 24V, 4-20 mA loop is grounded at one point only (i.e.

either at the input of the signal converter or at the input of the
indicator, or at the output of the Power Supply). Do not ground the
loop at more than one point.

3. Electrical Connection

3.1 Installation Location & Cable diameter

3.2 Connection to loop power supply

3.3.l Abbreviations

Abbreviation Stands for

Programming
via Fct. No....

Description
see Sect....

Electronic counter

Electro-mechanical counter

Pulse output

Pulses for Q =100% flow rate

0% flow rate

Full-scale range, 100%. flow rate.

EC

EMC

P

P

Q

Q

100%

0%

100%

-

-

1.3.X

1.3.2

1.1.2

1.1.2+1.3.2

3.2+3.3.4

3.2

6.1+6.2.9

6.2.9

6.1

6.1+6.2.9

• The current output is galvanically isolated from Pulse output P.

• Max. load at terminals 5/6

3.3 Outputs

3.3.2 Current (analog) output I

max. load in Kohms

U – 12 V

B

R =

B

< 1200 ohms

20mA