Technical data – KROHNE OPTITEMP TT 30 C-R EN User Manual
Page 47
TECHNICAL DATA
7
47
OPTITEMP TT 30 C/R
www.krohne.com
11/2012 - 4000753302 - MA OPTITEMP TT 30 C/R R02 en
Accuracies in °F
Input type
Temp. range
[°F]
Min. span
[°F]
Max. accuracy
Max. temp. influence
(dev. from ref. temp.
73°C)
RTD Pt100
-328...+1832
18
Max. of ±0.4°C or ±0.1% of
input span
1
Max. of ±0.01°F per °F or
±
0.006% of input span
per °F
2
RTD Pt1000
-328...+392
RTD PtX
3
-328...max.
input temp
(corresp.
to 2 kΩ)
RTD Ni100
-76...+482
RTD Ni120
-94...+572
RTD Ni1000
-148...+302
RTD Cu10
-328...+500
180
RTD Pt100
(α = 0.003902)
-328...+1832
18
RTD Pt100
(α = 0.003916)
T/C type B
+32...+3272
Corresp.
to 2 mV
±
20 μV or ±0.1% of input
span
4
±
0.01°F per °F or ±0.006%
of input span per °F
5
T/C type C
+14...+4172
T/C type E
-328...+1832
T/C type J
T/C type K
-328...+2462
T/C type L
-328...+1652
T/C type N
-454...+2372
T/C type R
-58...+3182
T/C type S
T/C type T
-328...+752
T/C type U
-328...+1112
T/C custom
-10...+500 mV
1 Linearity error is not included
2 If zero-deflection is > 100% of input span then you have to add 0.003% of input span per °F per 100% zero-deflection!
3 (10 ≤ X ≤ 1000)
4 Linearity error and CJC error is not included
5 Temp. influence CJC is not included; also note that if zero-deflection is > 100% of input span then you have to
add 0.003% of input span per °F per 100% zero-deflection!
.book Page 47 Friday, November 9, 2012 3:50 PM