Technical data – KROHNE OPTITEMP TT 30 C-R EN User Manual
Page 40
7
TECHNICAL DATA
40
OPTITEMP TT 30 C/R
www.krohne.com
11/2012 - 4000753302 - MA OPTITEMP TT 30 C/R R02 en
Input impedance
>10 MΩ
Maximum wire loop resistance
500 Ω (total loop)
Cold Junction Compensation
(CJC)
Internal or none (0°C)
Input - voltage
Input - voltage
Input - voltage
Input - voltage
Range
-10...+500 mV
Minimum span
2 mV
Customized linearization
Up to 9 points
Input impedance
>10 MΩ
Maximum wire loop resistance
500 Ω (total loop)
Output
Output
Output
Output
Output signal
4...20 mA, 20...4 mA or customized; temperature, resistance or voltage linear,
customized linearziation possible.
Span: 3.8...20.5 mA (measurement), 3.5...21.6 mA (failure)
Resolution: 5 μA
Update time
Approx. 1.5 seconds
Selectable dampening time
0 or 2 s (time constant)
NAMUR compliance
Current limitations and failure currents acc. to NAMUR NE 43
Configuration
Configuration
Configuration
Configuration
TempSoft
The PC configuration software, TempSoft, is a versatile and user-friendly tool for
transmitter configuration, loop check-up and sensor diagnostics.
It runs on Windows 2000, XP, Vista and Windows 7.
Approvals and certifications
CE
The device fulfils the statutory requirements of the EC directives.
The manufacturer certifies that these requirements have been met by applying the
CE marking.
Electromagnetic compatibility
According to EN 61326-1:2006
Ex approvals
Ex approvals
Ex approvals
Ex approvals
Non-Ex version
Without
ATEX
Intrinsically safe according to ATEX Directive 94/9/EC harmonized standards
In-head transmitter: EN 60079-0:2006, EN 60079-11:2007 and EN 60079-26:2007.
Rail-mount transmitter: EN 60079-0:2006, EN 60079-11:2007
Electromagnetic compatibility
Directive: 2004/108/EC
Harmonized standards according to EN 61326-1:2006
Vibration resistance
Acc. to IEC 60068-2-6; tested:
In-head: 10 g, 60...2000 Hz,
Rail: 5 g, 60...500 Hz
Shock resistance
Acc. to IEC 60068-2-31, tested for Drop and Topple, IEC 60068-2-31
NAMUR
Output limitations and fail currents according to NAMUR recommendations.
.book Page 40 Friday, November 9, 2012 3:50 PM