Operation, 4 configuring the rail-mount version – KROHNE OPTITEMP TT 10 C-R EN User Manual
Page 31
OPERATION
5
31
OPTITEMP TT 10 C/R
www.krohne.com
02/2013 - 4000752703 - MA OPTITEMP TT 10 C/R R03 en
5.4 Configuring the rail-mount version
• Take a screwdriver and press the tip between the housing and the cover as shown in the
previous drawing.
• Remove the cover.
DANGER!
To avoid electric shocks and destructions or damages of the device, assure that all electrical
connections comply with the instructions within the chapter "Electrical connections".
DANGER!
Only perform the configuration described in this section if no sensor is connected to the
transmitter. Otherwise you could damage the cable insulation during soldering, furthermore the
cables could handicap during soldering.
INFORMATION!
In the standard delivery condition the transmitters are not preset. Therefore you have to make a
complete solder pad configuration and adjust the potentiometers before using the transmitter
for the first time. Also note the following items:
•
The steps in this section show the complete configuration procedure in the correct order for
new transmitters. The manufacturer strongly recommends to follow this procedure,
otherwise the calibration may not be proper.
•
As an option the manufacturer offers preset transmitters according to the customer's order.
Only in this case the solder pad configuration and potentiometer adjustment is not necessary.
•
In some cases you have to change the solder pad configuration and perform a new
potentiometer adjustment of a transmitter that has already been in use. This may be the case
e.g. if you must define a new zero point or measuring span for a new application.
Step 1: removing the cover
.book Page 31 Tuesday, February 26, 2013 1:43 PM