UFM 3030

Part C Functional checks and service

7 Functional

checks

7.1

Test functions of the signal converter Function 2.1 to 2.5


The signal converter has several different test functions.

Display test, Function 2.01.01
Select function 2.01.01 as described in sect. 4.2 and 4.3.
Press Right arrow to start.
All segments in the 3 lines of the display are activated sequentially.
The test can be terminated by pressing the

↵

(return) key.


Test, current output I, Function 2.02.01
WARNING – this will affect the current output, and will no longer be process output!
A milliamp-meter must be connected between terminals V+ and I for this test, see the connection
diagrams in section 2.
Select Function 2.02.01 as described in sect. 4.2 and 4.3.
Select current value with

key:

0

mA

4

mA

12

Ma

20

mA

22

mA

The milliamp-meter indicates the current value selected. Press the

key to terminate the test and

display the actual value again

Test, frequency output F, Function 2.02.02
WARNING – this will affect the pulse/frequency output, and will no longer be process output!
An electronic Totalizer (EC) must be connected to terminal P.

for this test.

Select function 2.02.02 as described in sect. 4.2 and 4.3.
Select frequency value with

↵

key:

1

Hz

10

Hz

100

Hz

1000

Hz

2000

Hz

The Totalizer indicates the frequency value selected.
Press the

↵

key to terminate the test and display the actual value again.

Test of the analog inputs ( if applicable) 2.03.01 or 2.03.02
Connect a current source to terminal A1/GND or A2/GND.
Select Function 2.03.01 or 2.03.02.
Set the current source to a value 4..22mA.
Read the measured value from the display of the flow meter.
Press '<-' to terminate the test.
WARNING: a current >30mA will result in damage of the analog input. Make sure the current will
not exceed the 30mA.