KROHNE UFM 3030 Service EN User Manual
Page 24
24
UFM 3030
3.03.08
ERROR MSG
Display error messages
NO, YES
3.03.09 DATE
Display
date
NO, YES
3.03.10
AN INPUT
Display analog inputs
NO, YES
3.03.11
SIGN LEVEL
Display signal level
NO, YES
3.04.00
CURR OUTP
Submenu 3.04.00 Current output
3.04.01
FUNCTION
Function of current output
OFF (switched off)
ACT FLOW (actual flow)
CORR FLOW (corrected flow), see Function 3.02.02 and 3.02.08
through 3.02.11
F/R IND (forward/reverse indication of actual flow)
VOS (velocity of sound, range is defined in Function 3.01.11 and
3.01.12)
GAIN (sensor signal gain, range is 0 dBV through 100 dBV)
AN INP 1 (analog input 1)
AN INP 2 (analog input 2)
3.04.02
DIRECTION
Direction of current output
FORWARD (forward flow measurement)
BOTH (forward and reverse flow measurement indicating both in
the same range)
F/R SPEC (forward and reverse flow measurement indicated in
different range see Function 3.04.04)
3.04.03
RANGE
Range of current output
OTHER (user defined, see Function 3.04.04 through 3.04.06)
0-20/22 mA (0 pct - 100 pct / limit)
4-20/22 mA(0 pct - 100 pct / limit)
3.04.04
0 pct
Current value for 0% scale
Range: 0 through 16 mA
3.04.05
100 pct
Current value for 100% scale
Range: 4 through 20 mA
Value must be at least 4 mA greater than current value for 0%
scale
3.04.06 LIMIT
Limitation
of
current
value
Range: 20 through 22 mA
3.04.07
ERR INDIC
IERR = 3.6 MA
IERR = 21.5 MA
(only available for NAMUR devices)
3.05.00
PULSE OUTP
Submenu 3.05.00 Pulse output
3.05.01
FUNCTION
Function of pulse output
OFF (switched off)
ACT FLOW (actual flow)
CORR FLOW (corrected flow), see Function 3.02.01 and 3.02.08
through 3.02.11
F/R IND (forward/reverse flow indication)
VOS (velocity of sound, range defined in Function 3.01.11 and
3.01.12)
DIG OUTPUT (digital output, see Function 3.05.03)
BATCH OUTP (batch output indication, see Function 3.02.01)
GAIN (sensor signal gain, range is 0 dBV through 100 dBV)
AN INP 1 (analog input 1)
AN INP 2 (analog input 2)