4.03.03

CUTOFF OFF

Cutoff “off” value, see Function 3.01.07

4.04.00

ENERGY

Full scale value for heat energy rate (E) incorrect. The fullscale
value is compared with the maximum value that can be
measured and should meet the condition: Emax < E fullscale <
Emax/1000
The maximum value that can be measured is at maximum flow
and 200° C temperature difference.

4.04.01

HEAT FS

Full-scale value for 100 % heat energy rate, see Function 3.02.12

4.05.00

PULSE/VOS

Unit of pulse output for velocity of sound function incorrect
Ensure "PULSE RATE" is selected for "VOS"!

4.05.01

PULS FUNCT

Function of pulse output, see Function 3.05.01

4.05.02

PULSE OUTP

Unit of pulse output, see Function 3.05.07

4.06.00

VOS

Velocity of sound range incorrect:
Ensure condition MAX VOS

MIN VOS

1 m/s (3.3 feet/sec) is

met!

4.06.01

MIN VOS

Minimum velocity of sound, see Function 3.01.11

4.06.02

MAX VOS

Maximum velocity of sound, see Function 3.01.12

4.07.00

PULSE OUTP

Pulse output frequency value (f) incorrect. The max frequency is
calculated from the pulse/unit setting and the max value of the
measured value.
Ensure condition 1 pulse/hr

f

2000 pulse/s is met.

4.07.01

PULSE/UNIT

Pulse value for volume flow rate unit, see Function 3.05.09

4.07.02

PULSE/UNIT

Pulse value for heat power rate unit, see Function 3.05.10

4.08.00

PULS WIDTH

Pulse output pulse width incorrect
Ensure condition pulse width

0.5 x pulse period time is met.

4.08.01 PULS

WIDTH

Pulse width for frequencies

10 Hz, see Function 3.05.11

4.09.00

HART

Current output range for HART incorrect. If Hart is activated the
minimum possible current should be 4 mA.
Ensure condition CURR 0 pct

4 mA is met.

4.09.01 CURR

RANGE

Range of current output, see Function 3.04.03

4.09.02

CURR 0 pct

Current value for 0% scale, see Function 3.04.04

4.10.00

INP/OUTP

The digital input (C) and current output (I) are not allowed to be
switched on simultaneously. If the Profibus option is activated
only one of the following input/output functions can be used:
digital input (C), current output (I), pulse output (P). The current
output is deactivated by setting the function of current output to
off en setting the range of current output to 0-20mA.

4.10.01

INP FUNCT

Function of digital input, see Function 3.06.01

4.10.02

CURR FUNCT

Function of current output, see Function 3.04.01

4.10.03 CURR

RANGE

Range of current output, see Function 3.04.03

4.10.04

PULS FUNCT

Range of pulse output, see Function 3.05.01

4.13.00

EPROM

EPROM checksum error, reset device.