UFM
3030
27
4.03.03
CUTOFF OFF
Cutoff “off” value, see Function 3.01.07
4.04.00
ENERGY
Full scale value for heat energy rate (E) incorrect. The fullscale
value is compared with the maximum value that can be
measured and should meet the condition: Emax < E fullscale <
Emax/1000
The maximum value that can be measured is at maximum flow
and 200° C temperature difference.
4.04.01
HEAT FS
Full-scale value for 100 % heat energy rate, see Function 3.02.12
4.05.00
PULSE/VOS
Unit of pulse output for velocity of sound function incorrect
Ensure "PULSE RATE" is selected for "VOS"!
4.05.01
PULS FUNCT
Function of pulse output, see Function 3.05.01
4.05.02
PULSE OUTP
Unit of pulse output, see Function 3.05.07
4.06.00
VOS
Velocity of sound range incorrect:
Ensure condition MAX VOS
−
MIN VOS
≥
1 m/s (3.3 feet/sec) is
met!
4.06.01
MIN VOS
Minimum velocity of sound, see Function 3.01.11
4.06.02
MAX VOS
Maximum velocity of sound, see Function 3.01.12
4.07.00
PULSE OUTP
Pulse output frequency value (f) incorrect. The max frequency is
calculated from the pulse/unit setting and the max value of the
measured value.
Ensure condition 1 pulse/hr
≤
f
≤
2000 pulse/s is met.
4.07.01
PULSE/UNIT
Pulse value for volume flow rate unit, see Function 3.05.09
4.07.02
PULSE/UNIT
Pulse value for heat power rate unit, see Function 3.05.10
4.08.00
PULS WIDTH
Pulse output pulse width incorrect
Ensure condition pulse width
≤
0.5 x pulse period time is met.
4.08.01 PULS
WIDTH
Pulse width for frequencies
≤
10 Hz, see Function 3.05.11
4.09.00
HART
Current output range for HART incorrect. If Hart is activated the
minimum possible current should be 4 mA.
Ensure condition CURR 0 pct
≥
4 mA is met.
4.09.01 CURR
RANGE
Range of current output, see Function 3.04.03
4.09.02
CURR 0 pct
Current value for 0% scale, see Function 3.04.04
4.10.00
INP/OUTP
The digital input (C) and current output (I) are not allowed to be
switched on simultaneously. If the Profibus option is activated
only one of the following input/output functions can be used:
digital input (C), current output (I), pulse output (P). The current
output is deactivated by setting the function of current output to
off en setting the range of current output to 0-20mA.
4.10.01
INP FUNCT
Function of digital input, see Function 3.06.01
4.10.02
CURR FUNCT
Function of current output, see Function 3.04.01
4.10.03 CURR
RANGE
Range of current output, see Function 3.04.03
4.10.04
PULS FUNCT
Range of pulse output, see Function 3.05.01
4.13.00
EPROM
EPROM checksum error, reset device.