KROHNE UFM 3030 Service EN User Manual

Page 26

background image


26

UFM 3030

3.07.04

LOCATION

Tag name setting
Free settable tag for identification, maximum 10 characters.
Characters assignable to each place: A..Z / blank character / 0..9
Factory setting: KROHNE

3.07.05

UNIT TEXT

Text for user-defined unit
Definition: volume/time
Characters assignable to each place: A..Z / blank character / 0..9
Fraction bar "/" in 5th place is unalterable
Factory setting: XXXX/YYY

3.07.06

UNIT VOL

User-defined unit volume
Quantity of user-defined volume per m3.
Range: 10-5 through 107
Factory setting: 1

3.07.07

UNIT TIME

User-defined unit time
Amount of user-defined time in seconds
Range: 10-5 through 107
Factory setting: 1

3.07.08

RST ENABLE

Enable Totalizer reset
NO (Totalizer reset disabled)
YES (Totalizer reset enabled)

3.07.09

ERR LIMIT

Error limit in % of measured value for plausibility filter on the
sensor paths.

3.07.10

CNT DECR

Counter decrement setting for plausibility filter

3.07.11

CNT LIMIT

Counter limit for plausibility filter
When "0" is set, the plausibility filter will become inactive
Range: 0 through 1000
Factory setting: 0

3.09.00

COMMUNIC

Submenu 3.09.00 Communication

3.09.01 PROTOCOL Communication

protocol

OFF (no communication)
HART (HART)
PROFIB PA (PROFIBUS PA)

3.09.02

HART ADDR

HART address
Range: 00 through 16

3.09.03

PP/FF ADDR

PROFIBUS PA/FOUNDATION Fieldbus address
Range: 000 through 126

4.00.00

PARAM ERR

Main menu 4.00.00 Parameter error

4.01.00

FLOW VELOC Volume flow velocity (v) value incorrect. The flow speed is

calculated from the full scale volume flow and the meter size.
Ensure condition 0.5 m/s

v

20 m/s (1,5 to 66 feet/s) is met!

4.01.01

FULL SCALE

Full-scale value for 100% volume flow rate, see Function 3.01.01

4.01.02

METER SIZE

Meter size, see Function 3.01.08

4.02.00

CURR OUTP

Current output range incorrect. Setting for 100% is compared
with setting for 0%. Ensure condition 100 pct-0 pct

4 mA is met!

4.02.01

RANGE

Range of current output, see Function 3.04.03

4.02.02

0 pct

Current value for 0% scale, see Function 3.04.04

4.02.03

100 pct

Current value for 100% scale, see Function 3.04.05

4.03.00

LF CUTOFF

Low-flow cut-off range incorrect: If low flow cut-off is set to on,
the value for CUTOFF-OFF is compared with the value of
CUTOFF-ON on.
Ensure condition CUTOFF-OFF – CUTOFF-ON

1% is met!

4.03.01

LF CUTOFF

Low-flow cut-off, see Function 3.01.05

4.03.02

CUTOFF ON

Cutoff “on” value, see Function 3.01.06