KROHNE UFM 3030 Service EN User Manual
Page 26
26
UFM 3030
3.07.04
LOCATION
Tag name setting
Free settable tag for identification, maximum 10 characters.
Characters assignable to each place: A..Z / blank character / 0..9
Factory setting: KROHNE
3.07.05
UNIT TEXT
Text for user-defined unit
Definition: volume/time
Characters assignable to each place: A..Z / blank character / 0..9
Fraction bar "/" in 5th place is unalterable
Factory setting: XXXX/YYY
3.07.06
UNIT VOL
User-defined unit volume
Quantity of user-defined volume per m3.
Range: 10-5 through 107
Factory setting: 1
3.07.07
UNIT TIME
User-defined unit time
Amount of user-defined time in seconds
Range: 10-5 through 107
Factory setting: 1
3.07.08
RST ENABLE
Enable Totalizer reset
NO (Totalizer reset disabled)
YES (Totalizer reset enabled)
3.07.09
ERR LIMIT
Error limit in % of measured value for plausibility filter on the
sensor paths.
3.07.10
CNT DECR
Counter decrement setting for plausibility filter
3.07.11
CNT LIMIT
Counter limit for plausibility filter
When "0" is set, the plausibility filter will become inactive
Range: 0 through 1000
Factory setting: 0
3.09.00
COMMUNIC
Submenu 3.09.00 Communication
3.09.01 PROTOCOL Communication
protocol
OFF (no communication)
HART (HART)
PROFIB PA (PROFIBUS PA)
3.09.02
HART ADDR
HART address
Range: 00 through 16
3.09.03
PP/FF ADDR
PROFIBUS PA/FOUNDATION Fieldbus address
Range: 000 through 126
4.00.00
PARAM ERR
Main menu 4.00.00 Parameter error
4.01.00
FLOW VELOC Volume flow velocity (v) value incorrect. The flow speed is
calculated from the full scale volume flow and the meter size.
Ensure condition 0.5 m/s
≤
v
≤
20 m/s (1,5 to 66 feet/s) is met!
4.01.01
FULL SCALE
Full-scale value for 100% volume flow rate, see Function 3.01.01
4.01.02
METER SIZE
Meter size, see Function 3.01.08
4.02.00
CURR OUTP
Current output range incorrect. Setting for 100% is compared
with setting for 0%. Ensure condition 100 pct-0 pct
≥
4 mA is met!
4.02.01
RANGE
Range of current output, see Function 3.04.03
4.02.02
0 pct
Current value for 0% scale, see Function 3.04.04
4.02.03
100 pct
Current value for 100% scale, see Function 3.04.05
4.03.00
LF CUTOFF
Low-flow cut-off range incorrect: If low flow cut-off is set to on,
the value for CUTOFF-OFF is compared with the value of
CUTOFF-ON on.
Ensure condition CUTOFF-OFF – CUTOFF-ON
≥
1% is met!
4.03.01
LF CUTOFF
Low-flow cut-off, see Function 3.01.05
4.03.02
CUTOFF ON
Cutoff “on” value, see Function 3.01.06