Main menu 0.00.00 Error/Totalizer
This menu is accessible from the measuring mode by pressing the

key and entering "CODE 2"

(

↑→

).

Depending on the programming of Function 3.03.08 ERROR MSG, errors occurring during process
flow measurement are represented with flashing display lines and/or a compass field. Depending
on the programming of Function 3.03.07 CYCL DISP, the error messages alternate with the display
of the measured value(s) every 5 seconds, or they can be manually selected by pressing the

key.

1. Flashing line with number of errors that have occurred.
2. Flashing line with description of error message(s).
3. Flashing bar, indicating "new" errors, not yet
acknowledged.
4. Compass field, indicating measuring path error(s):

Indication of measuring path errors:

1, 2, 3: for measuring path 1,2 and 3, open or shorted sensor no measured value from
path.
4. Noise error, to much noise on measuring path(s). Flow meter functions outside
specification

The following list gives an alphabetical overview of error messages that can occur during process
flow measurement and what to do. The error messages only appear when Function 3.03.08
ERROR MSG is YES.

Error
message

Description of error message

What to do

ADC AN INP

Analog input internal error, A1 or A2

Switch off and on the flow meter. If the
error still exists, contact KROHNE
representative

COMMUNIC

Communication device internal error

Reset the error, wait for one minute. If
the error re-appears, contact KROHNE
representative

CURR > MAX

Current output overflow (> 22 mA)

Check flow velocity

DSP

Digital signal processor (DSP) internal
error

Only checked at power-up. Switch off
and on the flow meter. If the error still
exists, contact KROHNE representative

EE MENU

Menu parameters corrupted

Contact KROHNE representative

EE SERVICE

Service parameters internal error

Contact KROHNE representative

EMPTY PIPE

Measuring tube not completely filled,

Fill measuring tube completely