beautypg.com

KROHNE UFM 3030 Service EN User Manual

Page 25

background image


UFM

3030

25

3.05.02

DIRECTION

Direction of pulse output
FORWARD (forward flow measurement)
BOTH (forward and reverse flow measurement indicating both in
the same range)

3.05.03

DIG OUTPUT

Function of digital output
PATH ERR (measuring path error indication)
TOTAL ERR (Totalizer error indication)
ALL ERR (indication of all errors)
AN INP ERR (analog input error indication
OVERRANGE (overrange indication)
TRIP POINT (trips when actual flow (Q) goes over a set limit)

3.05.04

TRIP PNT 1

First trip point
Range: 0 through 120% of Q100%

3.05.05

TRIP PNT 2

Second trip point
Range: 0 through 120 % of Q100%

3.05.06

TIME CONST

Time constant of pulse output
25 ms
MASTER TC (see Function 3.01.04)

3.05.07

OUTPUT

Unit of pulse output (see Function 1.03.00)
PULSE FREQUENCY, pulses per unit time, see Function
3.05.08)
PULSE/UNIT, Totalizer pulse output, pulses per unit volume, see
Function 3.05.09

3.05.08

PULSE RATE

Pulse rate (frequency) value for 100 % scale
pulse/s, pulse/hr, pulse/min
Range: 1 pulse/hr through 2000 pulse/s

3.05.09

PULSE/UNIT

Pulse value per volume unit for totalisation
pulse/m3, pulse/l, pulse/US.Gal, pulse/bbl,
free user definable unit

3.05.10

PULSE/UNIT

Pulse value per heat energy unit for totalisation
Pulse/MJ, pulse/Gcal, pulse/Mcal, pulse/GJ

3.05.11 PULS

WIDTH

Pulse width for frequencies

10 Hz

25 ms, 50 ms, 100 ms, 200 ms, 500 ms

3.06.00

DIG INPUT

Submenu 3.06.00 Digital input

3.06.01

FUNCTION

Function of digital input
OFF (switched off)
RST TOTAL (reset display Totalizer)
RST ERROR (reset error messages)
FORCE ZERO (set outputs to minimum values)
BATCH (start batch)

3.07.00

USER DATA

Submenu 3.07.00 User data

3.07.01

LANGUAGE

Language for display texts
GB/USA (English)
D (German)
F (French)

3.07.02

ENTRY CODE

Entry code for setting mode
NO (entry with key only)
YES (entry with key and code 1, factory set on 9 x key, see
Function 3.07.03)

3.07.03

CODE 1

Code 1
Press any 9-keystroke combination and then press the same
combination again. Each keystroke is acknowledged by "

Ж

" in

the display. If both combinations are equal, "CODE OK" appears
and the new code can be stored, else "WRONG CODE" appears
and the desired code has to be entered again.