Resistive. This mode, as the name suggests, is
useful for resistive loads. Table 5 indicates how the
power supply responds to a command to go from
Output ON to OFF.
WARNING
Accessing the BOP after the output
is disabled in BATTERY mode is
hazardous because (1) high current
arcing is possible and (2) either the
external battery voltage, or the volt-
age (±Voltage Protection max) on
the BOP output terminals may be
dangerous. Therefore, for battery
and constant-voltage-type active
electronic loads it is recommended
that two properly rated external
switches be installed for safety:
one in series with the battery, and
one across the BOP output. After
the unit is set to OFF, first open the
switch in series with the battery,
then close the switch across the
BOP output to ensure safety before
handling BOP connections. When
connecting the battery, the switch
across the output should be
opened after the connections are
complete and then the switch in
series with the battery should be
closed. If the constant-voltage-type
active electronic load is adjusted to
zero before handling the power
supply-to-load connections, only
the switch across the BOP output
is required.
Battery. This mode is necessary for the power sup-
ply to function properly and safely with either battery
or constant-voltage-type active electronic loads.
This mode prevents the battery from discharging
during the OFF state. When the output is disabled
(set to OFF), the BOP will go to current mode, cur-
rent will be set to zero, with voltage protect and cur-
rent limit set to maximum. In this way the battery will
not be discharged while the output is OFF. For con-
stant-voltage-type active electronic loads this mode
stops energy flow during the OFF state. Table 5 indi-
cates how the power supply responds to a command
to go from Output ON to OFF.
CONFIGURE LOAD TYPE. To configure, press
%
from the power-up screen, then highlight Load
Type and press
!
to modify. (If a Password is
required, see Operator’s Manual for instructions.)
Highlight Active, Resistive or Battery (see explana-
tions and associated WARNINGS above) and press
$
to save. Then press
$
to save for power-up,
#
to abort, or
%
to apply the changes (without
saving for power-up) and exit. After configuring, the
new setting will be effective when the power supply
goes from output on to output off.
To restore factory default (Active), press
%
from
the power-up screen, highlight Load Type, press
!
, then press
@
to restore default. Press
$
to
save for power-up or
%
to exit
ENABLING/DISABLING OUTPUT POWER.
The BOP output can be disabled (OFF) or enabled
(ON) by toggling the STANDBY key in local mode.
The behavior of the unit when disabled depends on
the Load Type setting (See “Determining How the
Unit Responds when Output is OFF (Load Type)” on
page 12. and Table 5 for details).
IX — ADDITIONAL FEATURES.
The user is urged to refer to the Operator’s Manual
for full explanations all BOP 1KW features, includ-
ing:
• Passwords - three independent levels of
access
• Changing the Default Power up Settings
• Digital Remote Operation - using SCPI com-
mands via RS 232 or GPIB ports
• Analog Remote Operation - via Analog I/O
port
• Details about Protect Limits and Software-
controlled limits
• Storing/Recalling Power Supply Output Set-
tings
• Waveform Generation - Sine, Triangle,
±Ramp, Square and Level segments. Local
operation allows up to 16 waveforms, maxi-
mum 10 segments per waveform. Remote
operation allows 1 waveform, maximum of
126 segments, or using LIST commands.
• Operator Testing
• Calibration - via either local keypad or remote
SCPI commands
• Parallel/Series Configurations -increase cur-
rent capability, voltage capability, or both.