Caution – HT instruments HT157 User Manual
Page 25
HT155 - HT157
EN - 23
Line 1
Lines 2-7
Line 8
Line 9
Fig. 21: Screen of integrated values of Integral analysis
Line 1
It allows setting the current date/time and the measuring range in the
same way as for the instant value display
Lines 2- 7
These lines show in real time the values of the following quantities
integrated by the instrument: LAeqT, LCeqT, LZeqT, LAFmax, LAFmin,
LAFeqT, LASeqT, LAIeqT, LAE, E, Cpeak+, Cpeak-
Line 8
In this line, the left part shows the duration of the “Ts” measure set in the
programming menu (see § 4.6.2) and the measuring time “Tm”
5. Press the “
” key to start the integration measurement. The instrument automatically
shows the screen with the integrated values (see Fig. 21) and the “Run” message
appears on the display
6. When the instrument reaches the measuring time (Tm = Ts), the instrument stops the
integration measurement and the “End” message appears on the display. Data are
automatically saved in the instrument’s memory
7. By pressing the “
” key while a process is in progress, the instrument pauses
measurement and the “Pause” message appears on the display. In this situation:
By pressing the “
” key, the instrument stops measuring, shows the “Finish”
message on the display and automatically saves the data up to that instant
By pressing the “
” key, the measurement in progress is automatically cancelled
and the instrument goes back to the waiting condition, showing the “Ready”
message on the display
By pressing the “
” key again, the measurement starts again from the instant in
which it was stopped and the “Run” message appears on the display again
CAUTION
While a measurement is in progress (Run), it is not possible to access the
programming menu and it is necessary to first complete or stop it
The
“
” icon cannot be used for data integration measurement