5. CARRYING OUT MEASUREMENTS
The instrument is capable of carrying out different types of analyses of the sound pressure
level of a noise source:
1. Statistic analysis of noise
2. Spectrum analysis of noise with octave-band filter (1/1) (HT157 only)
3. Spectrum analysis of noise with third-octave band filter (1/3) (HT157 only)
For simplicity, the following descriptions are referred to the instrument model HT157
5.1.
STATISTIC ANALYSIS OF NOISE
5.1.1. INITIAL
OPERATIONS
1. Turn on the instrument by pressing the “
” key. The screen for the selection of the
analysis type is shown on the display for approximately 5s (see Fig. 10)
Fig. 10: Initial screens of statistic analysis
2. Use the keys “
” and “
” the option “Stat.Analyzer (00101)” and confirm with “
”.
The instrument shows the main menu of the statistic analysis, consisting of 6 icons on
the display
3. Select the relevant icon and press “
” to enter the specific menu. In detail:
Statistic the instrument measures the sound pressure level with weighting “A”,
with time constant Fast and equivalent noise level
24 Hours measurement of the sound pressure level in a fixed time period of 24
hours
Integral the instrument integrates the sound pressure level in the three frequency
weightings (A, C, Z) and in the three time constants (Slow, Fast, Impulse)
Data Section which contains the data of the recordings saved with the instrument
(see chapter 7)
Setup allows accessing the instrument’s programming menu (see § 4.6)
Calibrate section in which it is possible to calibrate the sensitivity level of the
microphone by using the provided portable calibrator HT151 (see § 6.1.3)
4. The indication next to the icon “
:XXX” in the right-hand top part of the display
indicates the number of the group of the datum which can be saved in the instrument’s
memory