6.3.2. Anomalous
situations
If the instrument does not
detect any power supply
at connector "230V 50Hz",
it displays the message
reported to the side.
LOW
10AE204 30.11.12
-----
----A
NO POWER SUPPLY
LEN: 10.0 SEC: 4.0
If the instrument detects
a resistance higher than
the limit set, it gives out
a long sound and the
following screen
appears on the display.
LOW
10AE204 30.11.12
Symbol
indicates a
resistance higher than the
limit set.
0.323
9.12A
LEN: 10.0 SEC: 4.0
Test current
The displayed results can be saved by pressing the SAVE key twice (see § 7.1).
If the instrument detects
a resistance higher than
the maximum
measurable value, it
gives out a long sound
and the following screen
appears on the display.
LOW
10AE204 30.11.12
Symbol ">" indicates a
resistance higher than the
maximum measurable
value.
>0.999
9.12A
LEN: 10.0 SEC: 4.0
Test current
The displayed results can be saved by pressing the SAVE key twice (see § 7.1)