EQUITEST
5071
EN - 12
"AUTO" mode results
At the end of the test, in
case the average value
of resistance Ravg
measured is lower than
5
, the instrument gives
out a double sound to
signal the positive
result of the test. The
following screen
appears on the display:
LOW
30.11.12
Average value of
resistance Ravg
1.05
R+ R-
1.07
1.03
219mA 219mA
AUTO 0.11
Value of test current
FUNC
CAL
The results can be saved by pressing the SAVE key twice (see § 7.1).
"RT+" or “RT-" mode results
During the test, an
audible beep sounds in
case the measured
resistance value exceeds
99.9.
If, at the end of the test,
the maximum RT+ or RT-
resistance value
measured is lower than
5
, the instrument gives
out a double sound to
signal the positive
result of the test. The
following screen appears
on the display:
LOW
30.11.12
Maximum value of
resistance measured
during test
1.07
219mA
RT+ 0.11
TIME: 10s
Test current
Duration of test
FUNC
CAL
The results can be saved by pressing the SAVE key twice (see § 7.1).