4-7. Interface Operation

This instrument equips RS-232 as standard device with a D-SUB 9 PIN
SHELL on the rear panel. Besides, the instrument also provides a GPIB
option device with a 24 PIN SHELL in blue. The configuration is
compliance with IEE488.
For further detailed operation, please refer to the Interface manual.

4-8. Input overload protection

The maximum allowable input is shown as table 4-1. Please proceed

the measurement accordingly.

Table 4-1:

FUNCTION RANGE MAXIMUN

INPUT

DCV

5V~1000V

1000Vdc or peak ac

ACV (AC+DC)

5V~1000V

1000V rms continuous &

10

7

V•Hz maximum

DCA,ACA(AC+DC)

500μA~2A

fuse protected: T2A 250V 1.5kA

breaking capacity

DC,AC20A(AC+DC)

20A

fuse protected: F20A 600V

100kA breaking capacity

DC,ACmV (AC+DC)

500mV

450V dc or ac peak

OHM

all ranges

450V dc or ac peak

CAPACITANCE

all ranges

450V dc or ac peak

RIPPLE

all ranges

450V dc or ac peak

WARNING: To avoid shock hazard and/or instrument
damage, do not apply input potentials that exceed the input
overload limits shown in table 4-1.

4-9. Input connections to common

WARNING: To avoid shock hazard and/or instrument
damage, do not connect the common input terminal to any
source of more than 500 volts DC or peak AC above earth
ground.