6. MEASUREMENT TECHNIQUES

6-1. dBm measurement technique

dBm is defined as above or below a 1mW reference. A voltage

measurement is converted to dBm using the following formula:

dBm=10*log

10

(1000*voltage value

2

/reference impedance.)

The reference impedance can be set.

For example, the reference impedance of 600Ω, 0.7746V will be

convert to 0 dBm.

6-2. True rms measurement

The true rms (root-mean-square) value of a waveform is equivalent to

dc value that causes the same amount of heat to be dissipated in a

resistor.

Since average-responding meters have been in use for so long, you

may have accumulated test or reference data based on them. Figure

6-1 illustrates the relationship between ac and dc components for

common waveforms, and compares readings for true rms meters and

average-responding meters. Figure 6-1 will help you convert between

the two measurement methods.

DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL


22

z Figure 6-1: Voltage Conversion