beautypg.com

GW Instek GDM-8246 User Manual User Manual

Page 7

background image

DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL


9

6. TRUE RMS AC OR AC+DC CURRENT

Accuracy

Between 2% of range and full range.

RANGE 20Hz-45Hz

45Hz-2kHz

2kHz-10kHz

10kHz-20kHz

500μA

5mA

50mA

500mA

1%+15 2%+15

2A

—————————

20A

1%+15 0.5%+15

—————————

Protection

Fuse protection 500μA,5mA,50mA,500mA,2A and 20A
ranges, 20A range input for 15 seconds max.

Crest Factor Range

3.0 at full scale.

The burden voltage is the same as the DC current.

When the input exceeds the full scale of the selected range, the display will appear
“—OL—” of over-range indication.

7. FREQUENCY MEASUREMENT AT ACA RANGE

RANGE FREQUENCY INPUT LEVEL (SINE WAVE)

ACCURACY

500μA 10Hz ~ 20kHz

≧90μA

5mA 10Hz

~

20kHz

≧0.9mA

50mA 10Hz

~

20kHz

≧9mA

500mA 10Hz

~

20kHz

≧90mA

2A 10Hz

~

2kHz

≧1A

20A 10Hz

~

2kHz

≧9A

0.05%+1

AC+DC measurement does not support AC+Hz function.

DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL


10

8. RESISTANCE

RANGE RESOLUTION

ACCURACY

500Ω 0.01Ω

0.1%+4

5kΩ 0.1Ω

50kΩ

500kΩ 10Ω

0.1%+2

5MΩ 100Ω

0.2%+2

20MΩ 1kΩ

0.3%+2

Open-circuit Voltage

3.2 volts maximum on 500Ω, 5kΩ 1.3 volts maximum on all

other ranges.

Protection

450V dc or peak ac continuous.

9. CAPACITANCE

RANGE: RESOLUTION

ACCURACY

5n *

0.001n

≧1nF: 2%+10

<1nF & ≧0.5nF: 2%+20

50n 0.01n

≧10nF: 2%+10

<10nF & ≧5nF: 2%+30

500n 0.1n

1n

50μ

10n

2%+4

*5n range tends to be interfered by the test lead’s impedance and position. For the
accuracy, please measure the range on the input terminal directly.
Protection

450V dc or peak ac continuous.

10. DIODE CHECK

Description

Display read forward voltage of diode.

Open Voltage

3.1V approx.

Maximum Forward
Voltage

1.5V

Protection

450V dc or peak ac continuous.