beautypg.com

Accuracy specifications(2) – GW Instek GDM-356 User Manual

Page 28

background image

28

Accuracy Specifications(2)

C. DC Current

Range Resolution

Overload
Protection

Accuracy

(2%+5)

200mA

20A

100

µ

A

10mA

(0.8%+1)

0.5A. 250V
fast type
fuse, 5x20
mm

Un-Fused

(1.5%+1)

Remarks:
l At 20A Range: For continuous measurement 10
seconds and interval not less than 15 minutes.
l Measurement voltage drop: Full range at 200mV.

D. AC Current

Range Resolution

Overload
Protection

Accuracy

(3%+5)

(1.8%+3)

200mA

20A

100

µ

A

10mA

(1%+3)

0.5A. 250V
fast type
fuse, 5x20
mm

Un-Fused

Remarks:
l At 20A Range: For continuous measurement 10
seconds and interval not less than 15 minutes.
l Measurement voltage drop: Full range at 200mV.
l Frequency reaponse: 40Hz~400Hz
l Display effective value of sine wave (mean value
response).

10

µ

A

20mA

20mA

10

µ

A

GDM-356