E. Measuring Resistance

(see figure 7)

Warning

To avoid damages to the Meter or to the devices under
test, disconnect circuit power and discharge all the
high-voltage capacitors before measuring resistance.

The resistance ranges are:
200

, 2k

, 20k

, 200k

, 2M

and 200M

.

To measure resistance, connect the Meter as follows:
1. Insert the red test lead into the V

terminal and

the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement
position in

range.

3. Connect the test leads across with the object being
measured.The measured value shows on the display.

Note
l The test leads can add 0.1

to 0.3

of error to the

resistance measurement.To obtain precision readings
in low-resistance, that is the range of 200

, short-

circuit the input terminals beforehand and record the
reading obtained (called this reading as X). (X) is
the additional resistance from the test lead.
Then use the equation:

Measurement Operation(7)