GDM-360/397/398/461 User Manual

67

Temperature Measurement (GDM-397 only)

Range

Resolution

Accuracy

Overload Protection

˚C

1˚C

(-40~-20˚C): -(8%+5)

1000Vdc / ac (Vpp)

(>-20~0˚C): ±(1.2%+4)
(>0~100˚C): ±(1.2%+3)
(>100~1000˚C): ±(2.5%+2)

˚F

1˚F

(-40~4˚F): -(8%+6)
(>4~32˚F): ±(1.2%+5)
(>32~212˚F): ±(1.2%+4)
(>212~1832˚F): ±(2.5%+3)

Thermocouple:
It is suitable to use K-type thermocouples. The included K-type thermocouple
can only be used to measure temperatures less than 230˚C.

Transistor hFE (GDM-398 only)

Range

Resolution

Remark

hFE

Ibo≈10μA; 1000β MAX