beautypg.com

GW Instek GDM-398 User Manual

Page 56

background image

GENERAL SPECIFICATIONS

54

GDM-360

Range

Resolution

Accuracy

Overload Protection

600μA

0.1μA

±(1.0%+3)

Fuse 1: F600mA H 1000V,φ6.35
x 31.8mm

6000μA

1μA

60mA

0.01mA

600mA

0.1mA

6A

0.001A

±(1.2%+5)

Fuse 2: F10A H 1000V,φ10.3 x
38.1mm

10A

0.01A

Remarks:

When ≤ 5A: Continuous measurement is allowed.

When > 5A: Continuous measurement for less than 10 seconds with intervals of
more than 15 minutes between measurements.