beautypg.com

2 calibration mode, 3 program mode, Calibration mode – Detcon 1000 User Manual

Page 17: Program mode

background image

Mdl 1000 H2S E-Chem / Inst. Air

Model 1000 H2S E-Chem w Instrument Air

Rev. 1.1

Page 13 of 28

5.2.2 Calibration Mode


Calibration mode allows for sensor zero and span adjustments. It uses an “AUTO SPAN” sequence with built
in diagnostics to ensure adequate sensor response and signal stability. “1-ZERO 2-SPAN

Zero Adjustment

Zero is set with no H

2

S target gases present. “AUTO ZERO


Span Adjustment

Span adjustment is performed with a target gas concentration of 10 ppm H

2

S in nitrogen. Span gas

concentrations other than 10 ppm may be used. “AUTO SPAN” See Table 3 Sample Flow Rate

5.2.3 Program Mode


Program Mode provides a program status menu (View Program Status) to check operational parameters.
Program Mode also allows for the adjustment of alarm set point levels, the calibration gas level setting,

View Program Status

View Program Status is a listing that allows the operator to view the gas, range, and software version number
of the program, as well as the current alarm settings, calibration gas level setting, offset value, RS-485 ID
number, heater voltage, and estimated remaining sensor life.

The following procedure is used to view the programming status of the sensor:

a) Enter the programming menu by holding the programming magnet stationary over “PGM 2” for 15

seconds. When the display changes to “VIEW PROG STATUS”, withdraw the magnet. Scroll through
the programming menu by momentarily waving the programming magnet over either “PGM 1” or “PGM
2”. The menu options are: View Program Status, Set Alarm 1 Level, Set Alarm 2 Level, and Set Cal
Level.

b) Scroll to the “VIEW PROG STATUS” listing and hold the programming magnet over “PGM 1” for 3

seconds. The display will automatically scroll, at five second intervals, through the following information
before returning back to the “VIEW PROG STATUS” listing.


¾ Gas type and software version number.
¾ Alarm set point level of alarm 1. The menu item appears as: “ALM1 SET @ ##PPM
¾ Alarm firing direction of alarm 1. The menu item appears as: “ALM1 ASCENDING” or descending.
¾ Alarm relay latch mode of alarm 1. The menu item appears as: “ALM1 NONLATCHING” or latching.
¾ Alarm relay energize state of alarm 1. The menu item appears as: “ALM1 DE-ENERGIZED” or

energized.

¾ Alarm set point level of alarm 2. The menu item appears as: “ALM2 SET @ ##PPM
¾ Alarm firing direction of alarm 2. The menu item appears as: “ALM2 ASCENDING” or descending.
¾ Alarm relay latch mode of alarm 2. The menu item appears as: “ALM2 LATCHING” or non-latching.
¾ Alarm relay energize state of alarm 2. The menu item appears as: “ALM2 DE-ENERGIZED” or

energized.

¾ Alarm relay latch mode of the fault alarm. The menu item appears as: “FLT NONLATCHING” or

latching.

¾ Alarm relay energize state of the fault alarm. The menu item appears as: “FLT ENERGIZED” or de-

energized.

¾ Calibration gas level setting. The menu item appears as: “CAL LEVEL @ ##PPM
¾ Identification of the RS-485 ID number setting. The menu item appears as: “485 ID SET @ ##
¾ Estimated remaining sensor life. The menu item appears as: “SENSOR LIFE 100%