If skilled technicians are available, replace just the UV lamp every 9-12 months.

3)

For abnormally high rates of signal decay, clean the UV lamp monthly, using a Lamp Cleaning Kit, and replace the

UV lamp every 9-12 months.

4)

For any proven cases where the zero baseline has drifted up, replace the detector cell.

3.8.5.2 PID Plug-In Sensor Maintenance Procedure

All piD Sensor Cells contain six user replacable parts.

Disassembly

1)

Power down the instrument and remove the sensor cell.

2)

Remove the f iltercap by applying a slight upward pressure with the tip of a screwdriver or an Exacto Blade just

below the hole in the cap and between the cap and the housing.

3)

With a f ine tipped tweezers, remove both of the Filter Media and set aside.

PI-600 Toxic Gas Sensors PG.26