beautypg.com

Detcon PI-600 User Manual

Page 23

background image

NOTE 2: Assuming that there are some residual VOC target gases in the background, you will require the use of the

zero air gas standard to perform a correct zero calibration starting from step a) below.

a) Apply a zero air standard at 200 cc/min for approximately 2-3 minutes then proceed through steps b) and c) with the

gas under continuous f low. After applying gas, tap your f ingertip over the exit port of the cal adapter for 10 -30

seconds to expedite the gas purging process.

b) Enter the calibration menu by holding the programming magnet over PGM 1 (see Figure #6) for 3 seconds until the

display reads “1-ZERO 2-SPAN”, then withdraw the magnet. Note that the Cal LED should now be illuminated.

c) Next, enter the zero menu by holding the magnet stationary over “PGM 1” for 3 seconds until the display reads:

“Setting Zero”, then withdraw the magnet. The sensor has now entered the auto zero mode. When it is complete

the display will read “ZERO COMPLETE” for 5 seconds and then return to the normal operations menu reading.

d) Remove the zero air gas and cal adapter and allow the sensor 3-5 minutes to rest on ambient air. If there are residual

active VOC gases in the area, then the sensor will read higher than 0.0 ppm. If this is the case, then you can use

the Zero Offset feature to correct for this residual background amount.

e) Set the Zero Offset value according to the concentration value found following the zero calibration procedure. See

section 3.8.1.1 below.

3.8.1.1 Using the Zero Offset Feature

If it is determined that there is a constant and non-negligible amount of residual active VOC gases in the background

air, the Zero Offset feature can be used to correct for this.

a) Observe the sensor’s concentration reading on air after a zero air calibration procedure. This represents the back-

ground VOC contribution that you will be offsetting.

b) Access the Zero Offset software feature by applying the magnet to PGM2 for 15 seconds. Then use the magnet to

momentarily pass over PGM1 and advance to the “Set Zero Offset” menu. Then apply the magnet to PGM1 for 3

seconds to access this menu. The menu should now read as “Zero Offset = X.X. Use PGM1 to increment this

number up to your desired offset level. When the correct offset is set, apply the magnet to PGM1 for 3 seconds to

accept the value. Then apply the magnet to PGM2 for 3 seconds to return to Normal Operation.

c) When done correctly, the unit should read 0.0 when back in Normal Operation.

3.8.2 Calibration Procedure - Span

3.8.2.1 Set Response Factor

All span calibrations are recommended to be done with a calibration standard consisting of isobutylene in an air back-

ground. If your target gas is different than the isobutylene span gas, you will be required to apply the correct Response

Factor. Look up the Response Factor for your target gas in the Table shown in Section 3.2.3.

a) Enter the programming menu by holding the magnet stationary over “PGM2” for 15 seconds until the display reads

“View Program Status”, then withdraw the magnet. At this point you can scroll through the programming menu by

momentarily waving the magnet over “PGM1” or “PGM2”. The menu options are: View Program Status, Set Cal Level,

Set Response Factor, and Set Zero Offset. Scroll to the “Set Response Factor” selection.

b) Select “Set Response Factor” by holding the magnet over “PGM1” for 3 seconds until the display reads “RespFactor

= x.xx”, then withdraw the magnet. Use the magnet to make an adjustment to “PGM1” to increase or “PGM2” to

decrease the displayed value until the value is equal to the desired “Response Factor” value from Section 3.2.3.

NOTE: If you have multiple target gases, then select the target gas with the highest Response Factor from the Table.

This provides for the safest and earliest warning levels.

NOTE: If you are span calibrating with the target gas, instead of isobutylene, then the response factor should be left at 1.0

3.8.2.2 Span Calibration

NOTE: Isobutylene is the recommended calibration gas for this sensor.

CAUTION: Verif ication of the correct calibration gas level setting and calibration span gas concentration is

required before “span” calibration. These two numbers must be equal.

PI-600 Toxic Gas Sensors PG.23