beautypg.com

Normal operation, Figure 14 ir-700 software flowchart, 3 normal operation – Detcon IR-700 User Manual

Page 18: Software flowchart

background image

Model IR-700

IR-700 Instruction Manual

Rev. 3.1

Page 14 of 40

Software Flowchart

dec

dec

Temp = XX C

mA Output = XX.XX

Ref Counts = XXXX

Voltage XX.X VDC

Sensor Life XXX%

Last Cal XX Days

Gas Factor = X.XX

Act Counts = XXXX

inc

PGM1/2 (M)

PGM2 (10)

Defaults Restored

PGM2 (S)

PGM1/2 (3)

PGM1 (S)

Restore Defaults

Auto Time-Out

##

Calibration Mode

(Auto Span)

AutoSpan @ XX

Range XXX

Serial ID XX

Gas Type

Auto Time-Out

Version X.XX

View Sensor Status

Model Type

PGM1/2 (3)

PGM1/2 (M)

inc

PGM1 (S)

PGM1/2 (3)

Auto Time-Out

PGM1/2 (M)

Set Serial ID

PGM1/2 (3)

AutoTime-out

PGM2 (S)

PGM1/2 (3)

PGM1/2 (M)

Set AutoSpan Level

##

PGM1 (3)

PGM2 (3)

Normal Operation

PGM1 (3)

PGM2 (10)

Calibration Mode

(Auto Zero)

inc - Increase
dec - Decrease

X, XX, XXX - numeric values

(S) - Momentary Swipe
(M) - Momentary hold of Magnet during text

scroll until the ">" appears, then release

(3) - 3 second hold from ">" prompt
(10) - 10 second hold from ">" prompt

Auto Time-out - 5 seconds

PGM1 - Program Switch Location #1
PGM2 - Program Switch Location #2

LEGEND:

Setting Gain

PGM1/2 (3)

Simulation

Auto Time-Out

Set Gas Factor

Auto Time-Out

Signal Output Check

Auto Time-Out

Set Sensor Gain

Auto Time-Out

PGM1 (S)

PGM1/2 (M)

PGM1/2 (3)

PGM1/2 (3)

inc

PGM1/2 (3)

PGM2 (S)

Type ##

Set Gas Type & Range

PGM1/2 (M)

dec

inc

dec

PGM1 (S)

PGM1/2 (10)

PGM1/2 (M)

PGM1/2 (3)

PGM1/2 (3)

PGM2 (S)

##

PGM1/2 (M)

PGM1 (S)

PGM1/2 (3)

inc

PGM2 (S)

Range ##

dec

Figure 14 IR-700 Software Flowchart

3.3 Normal Operation

In normal operation, the ITM Display continuously shows the current sensor reading, which will normally
appear as “0”. Once every 60 seconds the LED display will flash the sensor’s measurement units and gas type
(i.e. % LEL). If the sensor is actively experiencing any diagnostic faults, a “Fault Detected” message will
flash on the ITM display once every minute. When the unit is in “Fault Detected” mode, PGM1 or PGM2 can
be swiped to prompt the sensor to display the list of the active faults.

In normal operation, the 4-20mA current output corresponds with the present gas concentration and full-scale
range. The RS-485 Modbus™ serial output provides the current gas reading and fault status on a continuous
basis when polled.

Caution: Off-scale readings may indicate a flammable concentration of gas.