Setup menu – DeFelsko PosiTector UTG v.3.1 Full Manual User Manual

Page 16

background image

13

Reset

Reset (soft reset) restores factory settings and returns the
instrument to a known condition. The following occurs:

- All batches, stored measurements, images, and batch names

are erased.

- All calibration adjustments are cleared and returned to the

Gage’s factory calibration settings.

- Menu settings are returned to the following:

Memory = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Min Scan = OFF

A Scan = OFF
B Scan = OFF
Cal Lock = OFF
Bluetooth = OFF

Perform a more thorough Hard Reset by powering down the
instrument, waiting several seconds, then simultaneously holding
both the center and (+) buttons until the Reset symbol
appears. This returns the instrument to a known, “out-of-the-box”
condition. It performs the same function as a menu Reset with the
addition of:

Date and Time are not affected by either Reset.

Units = millimeter
Flip Display = Normal
Auto Sync = OFF
White on Black = OFF
Language = English

Battery Type = Alkaline
Backlight = Normal
Bluetooth Streaming = OFF
USB Drive= ON

- Bluetooth Pairing info is cleared.

- Menu settings are returned to the following:

Setup Menu

Setup Menu

NOTE:

Display = None
SE Mode = OFF
Smart Couple = OFF

Gage Info

Displays the model number & serial number, probe type & serial
number, PosiSoft.net registration key, the amount of remaining
memory for storage of readings, date and time, and software pack-
ages.

For security purposes, the registration key is required to add the
instrument to your PosiSoft.net account.