Calibration, Verification and Adjustment

Calibration, Verification and Adjustment

Three steps ensure best accuracy…

1.Calibration - typically done by the manufacturer or a qualified

lab. All probes include a Certificate of Calibration

2.Verification of Accuracy - as done by the user on known

reference standards such as calibration step blocks.

3.Adjustment - to a known thickness or sound velocity for the

material to be measured

Calibration

Calibration is the controlled and documented process of
measuring traceable calibration standards and verifying that the
results are within the stated accuracy of the Gage. Calibrations
are typically performed by the Gage manufacturer or by a certified
calibration laboratory in a controlled environment using a
documented process. The standards used in the calibration are
such that the combined uncertainties of the resultant
measurement are less than the stated accuracy of the Gage.

Written Calibration Procedures are available online at no charge
http://www.defelsko.com/downloads.htm

Verification

Gage accuracy can and should be verified using known reference
standards of the material to be tested.

Verification is an accuracy check performed by the user using
known reference standards. A successful verification requires the
Gage to read within the combined accuracy of the Gage and the
reference standards.

To guard against measuring with an improperly adjusted Gage,
verify the Gage at the beginning and the end of each work shift.
During the work shift, if the Gage is dropped or suspected of
giving erroneous readings, its accuracy should be re-verified.