beautypg.com

10 cal settings menu cal settings menu – DeFelsko PosiTector UTG v.3.1 Full Manual User Manual

Page 13

background image

10

Cal Settings Menu

Cal Settings Menu

The PosiTector UTG is factory calibrated. In order for it to take
accurate thickness measurements of a particular material it must
be set to the correct sound velocity for that material. Be aware that
material composition (and thus its sound velocity) can vary from
stated tables and even between lots from a manufacturer.
Adjustment to a sample of known thickness of the material to be
measured ensures that the Gage is adjusted as close as possible
to the sound velocity of that specific material. Samples should be
flat, smooth and as thick as the maximum expected thickness of
the piece to be tested.

The PosiTector UTG allows four simple adjustment choices. All
four methods are based on the simple premise of adjusting the
sound velocity. The first three adjustment methods make 1-point
calibration adjustments to optimize the linearity of the PosiTector
UTG
over small ranges. The fourth method makes a 2-point
calibration adjustment to allow for greater accuracy over a large
range.

Zero

PosiTector UTG probes can be "zeroed" to compensate for
temperature and wear. This action is not required when operating
in multiple echo mode (UTG M probes only) except when the
instrument is new and after a Reset (pg. 13). However, it ensures
best accuracy when operating in SE (single-echo) mode.
Therefore, it is a good idea to "zero" the probe before each work
shift and when the temperature of the part changes. During
zeroing, it does not matter what the current velocity setting is.

1. Make sure the instrument is on and the probe is wiped clean.

2. Apply a single drop of couplant onto the zero plate located on

the underside of the probe connector. DO NOT apply couplant
directly onto the probe face.

3. Select the Zero menu option and follow the on-screen prompts.

Calibration Adjustment