2. MEASURING MODE (POWER KEY)

2-1. Basic Display for Measuring Mode and Explanation

Display Contents

Status

Gas Name and Base Gas

Standard measuring mode.

CAUTION

CHECK AIR CAL.

Zero drifting might be occurring. Let the monitor sample fresh

air and perform AIR CAL.

CAUTION

ABNORMAL TEMP.

Internal thermistor detects an abnormal temperature.

CAUTION

LOW BATT.

Battery voltage is low. Replace the batteries.

CAUTION

LOW CONTRASUT

The interferometer sensor is getting contaminated. Before

measuring, the sensor should be replaced or repaired.

CAUTION

LOW BRIGHTNESS

The brightness is getting low. Before measuring, the sensor

should be replaced or repaired.

Caution

When the message “CAUTION ~ ” appears on the LCD, make an appropriate adjustment as soon as

possible.