TYPE-56 [FOR CALORIE MEASUREMENT]

Selectable Measuring Gas and Range

Measuring Gas

(Display Indication)

Measuring

Range

(Minimum

Digit)

AIR CAL

Value

Drift

Amount

Spec. No.

1

Propane in AIR

(PROPANE in AIR)

0~100 MJ/m3

Gross (0.1)

0.0

MJ/m3

0.6

MJ/m3

SPE- 1423

2

Propane in AIR

(PROPANE in AIR)

0~100 MJ/m3

Net (0.1)

0.0

MJ/m3

0.6

MJ/m3

SPE- 1424

3

Butane in AIR

(BUTANE in AIR)

0~100 MJ/m3

Gross (0.1)

0.0

MJ/m3

0.6

MJ/m3

SPE- 1425

4

Butane in AIR

(BUTANE in AIR)

0~100MJ/m3

Net (0.1)

0.0

MJ/M3

0.6

MJ/m3

SPE- 1426

5

Propane in AIR

(PROPANE in AIR)

0~20000

Gross

(20)

0

140

SPE- 1450

6

Propane in AIR

(PROPANE in AIR)

0~20000 Net

(20)

0

140

SPE- 1451

7

Butane in AIR

(BUTANE in AIR)

0~20000

Gross

(20)

0

140

SPE- 1452

8

Butane in AIR

(BUTANE in AIR)

0~20000 Net

(20)

0

140

SPE- 1453

Interferometer Type: Type 30MM

Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount

AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.

Drift Amount: The amount of indication drifting when temperature changes ±10ºC.

Special Caution:

1) This instrument is not explosion proof. This instrument measures combustible gas

around lower explosive limits in this specification. Take enough sample for the

measurement.

2) The measuring range 0-20000 is equivalent to old unit (kcal). Please read it as a

reference value.

3) Calorie amount for pure Propane, pure normal-Butane, and pure iso-Butane for this

instrument is defined and indicated as following the table (Ref. JISK2301-1992-9.)