Quit measuring and turn off instrument – COSA Xentaur SF6 Purity Pro User Manual

Page 35

background image

35

The higher that the flow rate of meter “A” is set, the sooner its indication will be stable. Also, since

the instrument has air in the internal piping, regulator, or sampling tube, we recommend adjusting

the flow monitor “A” so that it is more than 0.8 L/min every first turn when measuring.

SECTION 5. OTHER OPERATION

5-1. Incase of checking zero or suspending measuring, turn the change over cock (measuring

gas/air) sideways.

5-2. It is enough for further measuring to set flow meter “A” 0.5 L/min

*The unit can also accept the flow rate being set still 0.8 L/min.

Quit Measuring and Turn Off Instrument

SECTION 1. APPLY FRESH AIR

1-1. It is necessary for the unit to apply fresh air before turning off. Turn the change over the cock

(measuring gas/air) sideways to apply fresh air.

1-2. Press the power button until the LCD display goes out. The buzzer will beep three times while

pressing the power button.

SECTION 2. CLOSE THE CARRYING CASE

Press the stopper. Then put on the retaining bar once the box is closed.